A UNIVERSAL NITRIC-OXIDE DELIVERY SYSTEM

被引:17
作者
YOUNG, JD
机构
[1] Nuffield Department of Anaesthetics, Radcliffe Infirmary, Oxford OX2 6HE, Woodstock Road
关键词
PHARMACOLOGY; NITRIC OXIDE; DRUG DELIVERY;
D O I
10.1093/bja/73.5.700
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
A nitric oxide delivery system for adult ventilators has been developed; this is designed to maintain a constant inspired nitric oxide concentration of 16 volumes per million (vpm) in spite of changes in minute and tidal volumes. In a laboratory test the system maintained tracer gas concentrations equivalent to nitric oxide 13.8-18 vpm for tidal volumes of 500, 700 and 900 ml, ventilator rates of 10, 15 and 20 b.p.m., peak inspiratory flow rates of 30, 40 and 50 litre min(-1) and square, sine and decelerating ramp flow profiles.
引用
收藏
页码:700 / 702
页数:3
相关论文
共 4 条
[1]   ENHANCED PNEUMONIA RESOLUTION BY INHALATION OF NITRIC-OXIDE [J].
BLOMQVIST, H ;
WICKERTS, CJ ;
ANDREEN, M ;
ULLBERG, U ;
ORTQVIST, A ;
FROSTELL, C .
ACTA ANAESTHESIOLOGICA SCANDINAVICA, 1993, 37 (01) :110-114
[2]   LOW-DOSE INHALATIONAL NITRIC-OXIDE IN PERSISTENT PULMONARY-HYPERTENSION OF THE NEWBORN [J].
KINSELLA, JP ;
NEISH, SR ;
SHAFFER, E ;
ABMAN, SH .
LANCET, 1992, 340 (8823) :819-820
[3]   INHALED NITRIC-OXIDE IN PERSISTENT PULMONARY-HYPERTENSION OF THE NEWBORN [J].
ROBERTS, JD ;
POLANER, DM ;
LANG, P ;
ZAPOL, WM .
LANCET, 1992, 340 (8823) :818-819
[4]   INHALED NITRIC-OXIDE FOR THE ADULT RESPIRATORY-DISTRESS SYNDROME [J].
ROSSAINT, R ;
FALKE, KJ ;
LOPEZ, F ;
SLAMA, K ;
PISON, U ;
ZAPOL, WM .