DOPPLER-ECHOCARDIOGRAPHIC COMPARISON OF SMALL (19MM) BILEAFLET AND PERICARDIAL HEART-VALVE PROSTHESES IN AORTIC POSITION

被引:12
作者
GONZALEZJUANATEY, JR [1 ]
ACUNA, JMG [1 ]
AMARO, A [1 ]
CASTELO, V [1 ]
PEDREIRA, M [1 ]
BENGOECHEA, JG [1 ]
GIL, M [1 ]
机构
[1] HOSP XERAL DE GALICIA,SCH MED,DEPT CARDIAC SURG,E-15705 SANTIAGO,SPAIN
来源
SCANDINAVIAN JOURNAL OF THORACIC AND CARDIOVASCULAR SURGERY | 1995年 / 29卷 / 01期
关键词
PERICARDIAL BIOPROSTHESES; BILEAFLET PROSTHESES; 19MM AORTIC VALVE PROSTHESES; HEMODYNAMICS;
D O I
10.3109/14017439509107198
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
The resting haemodynamics of five types of small (19 mm) aortic valve prosthesis (2 bileaflet, 3 pericardial) were evaluated with Doppler echocardiography in 43 patients. Two received St Jude Medical and six CarboMedics bileaflet valves and 35 were given bioprostheses-16 Ionescu-Shiley, four Mitroflow and 15 Labcor-Santiago. No significant differences in peak or mean transvalvular pressure drop or in effective valve area were found between the bileaflet and the pericardial valves or among the three types of bioprosthesis. All but one of the bileaflet prostheses showed a characteristic regurgitation pattern, with two lateral and one central jet, and 16 (46%) of the bioprostheses showed central regurgitation, but in no case were these jets haemodynamically significant. Thus the 19 mm bileaflet and the studied pericardial prostheses all have satisfactory resting haemodynamics, and all are suitable for implanting in small aortic roots.
引用
收藏
页码:29 / 35
页数:7
相关论文
共 33 条
[1]   PROGNOSTIC FACTORS IN AORTIC-VALVE REPLACEMENT ASSOCIATED WITH NARROW AORTIC ROOTS - AN ANALYSIS USING THE PROPORTIONAL HAZARD MODEL [J].
ABDELNOOR, M ;
HALL, KV ;
NITTERHAUGE, S ;
LINDBERG, H ;
OVRUM, E .
SCANDINAVIAN JOURNAL OF THORACIC AND CARDIOVASCULAR SURGERY, 1986, 20 (03) :227-232
[2]   DOPPLER AND ECHOCARDIOGRAPHIC FEATURES OF NORMAL AND DYSFUNCTIONING BIOPROSTHETIC VALVES [J].
ALAM, M ;
ROSMAN, HS ;
LAKIER, JB ;
KEMP, S ;
KHAJA, F ;
HAUTAMAKI, K ;
MAGILLIGAN, DJ ;
STEIN, PD .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1987, 10 (04) :851-858
[3]   DISCREPANCIES BETWEEN DOPPLER AND CATHETER GRADIENTS IN AORTIC PROSTHETIC VALVES INVITRO - A MANIFESTATION OF LOCALIZED GRADIENTS AND PRESSURE RECOVERY [J].
BAUMGARTNER, H ;
KHAN, S ;
DEROBERTIS, M ;
CZER, L ;
MAURER, G .
CIRCULATION, 1990, 82 (04) :1467-1475
[4]  
BOJAR RM, 1989, J THORAC CARDIOV SUR, V98, P1087
[5]  
BOVE EL, 1985, J THORAC CARDIOV SUR, V90, P750
[6]   COMPARATIVE-STUDY OF THE HYDRODYNAMIC FUNCTION OF THE CARBOMEDICS VALVE [J].
BUTTERFIELD, M ;
FISHER, J ;
DAVIES, GA ;
SPYT, TJ .
ANNALS OF THORACIC SURGERY, 1991, 52 (04) :815-820
[7]   DOPPLER ECHOCARDIOGRAPHIC ASSESSMENT OF THE ST-JUDE MEDICAL PROSTHETIC VALVE IN THE AORTIC POSITION USING THE CONTINUITY EQUATION [J].
CHAFIZADEH, ER ;
ZOGHBI, WA .
CIRCULATION, 1991, 83 (01) :213-223
[8]   ECHOCARDIOGRAPHIC DESCRIPTION OF THE CARBOMEDICS BILEAFLET PROSTHETIC HEART-VALVE [J].
CHAMBERS, J ;
CROSS, J ;
DEVERALL, P ;
SOWTON, E .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1993, 21 (02) :398-405
[9]  
CHAMBERS J, 1992, J THORAC CARDIOV SUR, V104, P495
[10]  
COSGROVE DM, 1985, J THORAC CARDIOV SUR, V89, P358