AGREEMENT BETWEEN NONINVASIVE OXIMETRIC VALUES FOR OXYGEN-SATURATION

被引:26
作者
CAHAN, C [1 ]
DECKER, MJ [1 ]
HOEKJE, PL [1 ]
STROHL, KP [1 ]
机构
[1] CASE WESTERN RESERVE UNIV,DEPT MED,CLEVELAND,OH 44106
关键词
D O I
10.1378/chest.97.4.814
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
We made an assessment of five pulse oximeters in regard to their ability to replace the HP ear oximeter as a noninvasive measurement of SaO2. Trials were performed during isocapnic progressive hypoxia (SaO2 range, 99 to 70 percent) in 22 white and six black subjects. Comparisons between values of SaO2 by oximetry were analyzed by comparing the difference of values by the two methods against their mean. The difference between pulse oximeters and the HP was 2.6 ± 10.3 percent in all subjects. Results in whites were 1.9 ± 10.2 percent and in blacks were 5.1 ± 9.2 percent. The distribution of differences between pulse oximeters and the HP were larger below 80 percent than above 85 percent. We conclude that pulse oximeters give higher values than the HP, a tendency which is more pronounced in black than in white subjects. While the limits of agreement are better at saturations above 85 percent, the 95 percent confidence limits of agreement between pulse oximeters and the HP are rather large (± 10 percent) and unacceptable for assuming that pulse oximeters will provide the same values as found in clinical studies using the HP.
引用
收藏
页码:814 / 819
页数:6
相关论文
共 17 条
[1]  
ANDERSON JV, 1984, CLIN RES, V32, pA805
[2]   STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1986, 1 (8476) :307-310
[3]  
BROWN DP, 1987, 9TH P C IEEE ENG MED
[4]  
CHAUDHARY BA, 1978, AM REV RESPIR DIS, V117, P173
[5]  
DOUGLAS NJ, 1979, AM REV RESPIR DIS, V119, P311
[6]   CONTINUOUS INVIVO MONITORING OF ARTERIAL OXYGENATION IN CHRONIC OBSTRUCTIVE LUNG-DISEASE [J].
FLICK, MR ;
BLOCK, AJ .
ANNALS OF INTERNAL MEDICINE, 1977, 86 (06) :725-730
[7]  
MERRICK EB, 1976, HEWLETT-PACKARD J, P2
[8]  
MIHM FG, 1983, ANESTHESIOLOGY, V59, pA136
[9]  
POLOGE JA, INT ANESTHESIOL CLIN, V25, P137
[10]   THE ACCURACY AND RESPONSE CHARACTERISTICS OF A SIMPLIFIED EAR OXIMETER [J].
REBUCK, AS ;
CHAPMAN, KR ;
DURZO, A .
CHEST, 1983, 83 (06) :860-864