OBSERVER PERFORMANCE COMPARISON OF DIGITAL RADIOGRAPH SYSTEMS FOR STEREOTAXIC BREAST NEEDLE-BIOPSY

被引:9
作者
KRUPINSKI, EA [1 ]
ROEHRIG, H [1 ]
YU, T [1 ]
机构
[1] UNIV ARIZONA,ARIZONA HLTH SCI CTR,CTR OPT SCI,TUCSON,AZ 85724
关键词
CHARGE-COUPLED DEVICE DETECTORS; DIGITAL RADIOGRAPH CAMERAS; LANEX REGULAR; MIN-R REGULAR; OBSERVER PERFORMANCE; MAMMOGRAPHY;
D O I
10.1016/S1076-6332(05)80144-8
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Rationale and Objectives. Two digital radiograph systems for stereotactic mammography, one using a lens to couple a Lanex Regular screen to a back-illuminated charge-coupled device (CCD) and one using a fiber-optic taper to couple a Min-R Regular-type screen to a front-illuminated CCD, were evaluated with respect to observer performance. Methods. A contrast-detail phantom was imaged in a variety of equivalent exposure conditions on both systems. Six observers viewed images on a video monitor and recorded which objects were detected. Results. Performance (percent correct detections) with the lens-coupled system using the Lanex Regular screen was significantly higher than with the fiber-optic-coupled system using the Min-R Regular-type screen, Conclusion. Differences in absorption efficiencies of phosphors used, as well as differences in design of the two cameras, can explain differences in observer detection performance.
引用
收藏
页码:116 / 122
页数:7
相关论文
共 18 条
[1]   INTERVENTIONAL PROCEDURES IN DISEASES OF THE BREAST - NEEDLE-BIOPSY, PNEUMOCYSTOGRAPHY, AND GALACTOGRAPHY [J].
FAJARDO, LL ;
JACKSON, VP ;
HUNTER, TB .
AMERICAN JOURNAL OF ROENTGENOLOGY, 1992, 158 (06) :1231-1238
[2]   MAMMOGRAPHY-GUIDED STEREOTAXIC FINE-NEEDLE ASPIRATION CYTOLOGY OF NONPALPABLE BREAST-LESIONS - PROSPECTIVE COMPARISON WITH SURGICAL BIOPSY RESULTS [J].
FAJARDO, LL ;
DAVIS, JR ;
WIENS, JL ;
TREGO, DC .
AMERICAN JOURNAL OF ROENTGENOLOGY, 1990, 155 (05) :977-981
[3]   TEST PATTERN FOR VIDEO DISPLAYS AND HARD-COPY CAMERAS [J].
GRAY, JE ;
LISK, KG ;
HADDICK, DH ;
HARSHBARGER, JH ;
OOSTERHOF, A ;
SCHWENKER, R .
RADIOLOGY, 1985, 154 (02) :519-527
[4]  
HERRON JM, 1984, P SOC PHOTO-OPT INST, V486, P141, DOI 10.1117/12.943211
[5]  
HERRON JM, 1990, P SOC PHOTO-OPT INS, V1231, P374, DOI 10.1117/12.18813
[6]   SPATIAL-RESOLUTION IN DIGITAL MAMMOGRAPHY [J].
KARSSEMEIJER, N ;
FRIELING, JTM ;
HENDRIKS, JHCL .
INVESTIGATIVE RADIOLOGY, 1993, 28 (05) :413-419
[7]  
Kruger R A, 1984, BASIC CONCEPTS DIGIT
[8]  
LISK KG, 1984, P SOC PHOTO-OPT INST, V486, P79, DOI 10.1117/12.943201
[9]   ANALYSIS OF THE SPATIAL-FREQUENCY-DEPENDENT DQE OF OPTICALLY COUPLED DIGITAL MAMMOGRAPHY DETECTORS [J].
MAIDMENT, ADA ;
YAFFE, MJ .
MEDICAL PHYSICS, 1994, 21 (06) :721-729
[10]  