DETECTION OF SUBTLE MICROCALCIFICATIONS - COMPARISON OF COMPUTED RADIOGRAPHY AND SCREEN-FILM MAMMOGRAPHY

被引:36
作者
HIGASHIDA, Y
MORIBE, N
MORITA, K
KATSUDA, N
HATEMURA, M
TAKADA, T
TAKAHASHI, M
YAMASHITA, J
机构
[1] KUMAMOTO UNIV HOSP,DEPT RADIOL,KUMAMOTO,JAPAN
[2] KUMAMOTO UNIV,SCH MED,DEPT RADIOL,KUMAMOTO 862,JAPAN
[3] KUMAMOTO UNIV,SCH MED,DEPT SURG 2,KUMAMOTO 862,JAPAN
关键词
BREAST; CALCIFICATION; BREAST RADIOGRAPHY; RADIOGRAPHY; COMPARATIVE STUDIES; DIGITAL; STORAGE PHOSPHOR;
D O I
10.1148/radiology.183.2.1561354
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Observer performance tests were conducted to compare the effects of image-processing technique (unsharp mask filtering) on the diagnostic accuracy of computed radiography (CR) with storage phosphors in the detection of microcalcifications. Comparison of detectability of microcalcifications with CR and with screen-film mammography was also performed. Clusters of microcalcifications (125-250-mu-m in diameter) were randomly superimposed on human breast specimens. Observer performance tests were carried out with receiver operating characteristic (ROC) analysis. The area under the ROC curve and the cumulative true-positive-localization fraction were both used as indexes of performance. Observer performance experiments with nine observers indicated that the two types of screen-film images used provided higher detectability than the CR images. The detectability of microcalcifications on the CR images improved slightly with use of the unsharp masking technique. However, no statistically significant difference was found between processed and unprocessed CR images, and detectability still did not reach the level achieved with the screen-film images.
引用
收藏
页码:483 / 486
页数:4
相关论文
共 25 条
  • [1] DIGITAL MAMMOGRAPHY - ROC STUDIES OF THE EFFECTS OF PIXEL SIZE AND UNSHARP-MASK FILTERING ON THE DETECTION OF SUBTLE MICROCALCIFICATIONS
    CHAN, HP
    VYBORNY, CJ
    MACMAHON, H
    METZ, CE
    DOI, K
    SICKLES, EA
    [J]. INVESTIGATIVE RADIOLOGY, 1987, 22 (07) : 581 - 589
  • [2] EXCRETORY UROGRAPHY USING COMPUTED RADIOGRAPHY
    FAJARDO, LL
    HILLMAN, BJ
    HUNTER, TB
    CLAYPOOL, HR
    WESTERMAN, BR
    MOCKBEE, B
    [J]. RADIOLOGY, 1987, 162 (02) : 345 - 351
  • [3] DETECTION OF PNEUMOTHORAX - COMPARISON OF DIGITAL AND CONVENTIONAL CHEST IMAGING
    FAJARDO, LL
    HILLMAN, BJ
    POND, GD
    CARMODY, RF
    JOHNSON, JE
    FERRELL, WR
    [J]. AMERICAN JOURNAL OF ROENTGENOLOGY, 1989, 152 (03) : 475 - 480
  • [4] INVESTIGATION OF BASIC IMAGING PROPERTIES IN DIGITAL RADIOGRAPHY .6. MTFS OF II-TV DIGITAL IMAGING-SYSTEMS
    FUJITA, H
    DOI, K
    GIGER, ML
    [J]. MEDICAL PHYSICS, 1985, 12 (06) : 713 - 720
  • [5] BASIC IMAGING PROPERTIES OF A COMPUTED RADIOGRAPHIC SYSTEM WITH PHOTOSTIMULABLE PHOSPHORS
    FUJITA, H
    UEDA, K
    MORISHITA, J
    FUJIKAWA, T
    OHTSUKA, A
    SAI, T
    [J]. MEDICAL PHYSICS, 1989, 16 (01) : 52 - 59
  • [6] HIGASHIDA Y, 1989, Nippon Acta Radiologica, V49, P1411
  • [7] HIGH-SPEED, SINGLE-SCREEN SINGLE-EMULSION FILM SYSTEMS - BASIC IMAGING PROPERTIES AND PRELIMINARY CLINICAL-APPLICATIONS
    HIGASHIDA, Y
    FRANK, PH
    DOI, K
    [J]. RADIOLOGY, 1983, 149 (02) : 571 - 577
  • [8] IMAGING PERFORMANCE OF A DIGITAL STORAGE PHOSPHOR SYSTEM
    HILLEN, W
    SCHIEBEL, U
    ZAENGEL, T
    [J]. MEDICAL PHYSICS, 1987, 14 (05) : 744 - 751
  • [9] KEHLER M, 1989, ACTA RADIOL, V30, P581
  • [10] DIGITAL MAMMOGRAPHY - A COMPARISON OF 2 DIGITIZATION METHODS
    KIMMESMITH, C
    BASSETT, LW
    GOLD, RH
    GORMLEY, L
    [J]. INVESTIGATIVE RADIOLOGY, 1989, 24 (11) : 869 - 875