ABLATION OF THE ATRIOVENTRICULAR JUNCTION WITH RADIOFREQUENCY ENERGY USING A NEW ELECTRODE CATHETER

被引:81
作者
LANGBERG, JJ
CHIN, M
SCHAMP, DJ
LEE, MA
GOLDBERGER, J
PEDERSON, DN
OEFF, M
LESH, MD
GRIFFIN, JC
SCHEINMAN, MM
机构
[1] UNIV CALIF SAN FRANCISCO,DEPT MED,SAN FRANCISCO,CA 94143
[2] UNIV CALIF SAN FRANCISCO,CARDIOVASC RES INST,SAN FRANCISCO,CA 94143
关键词
D O I
10.1016/0002-9149(91)90436-O
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Percutaneous catheter ablation using radiofrequency energy can be used to interrupt atrioventricular (AV) conduction in patients with supraventricular tachycardia refractory to drugs. Results of radiofrequency ablation of the AV junction using a custom-designed catheter with a large, 3-mm-long distal electrode, 2-mm interelectrode spacing, and a shaft with increased torsional rigidity were compared with those using a standard quadripolar electrode catheter (Bard EP). An electrocoagulator (Microvasive Bicap 4005) supplied unmodulated radiofrequency current at 550 kHz, which was applied between the distal electrode of the ablation catheter and a large skin electrode. With use of the modified catheter, 12 of 13 patients (92%) had persistent complete AV block induced with 7 +/- 5 applications of 18 +/- 6 W of radiofrequency power. In contrast, complete AV block was produced in only 9 of 18 (50%) historical control patients treated with the standard catheter, despite a similar number of applications (7 +/- 5) and power output (16 +/- 4 W). A rise in impedance, due to desiccation of tissue and coagulum formation, occurred earlier (28 +/- 18 vs 52 +/- 24 seconds, p < 0.001) and more frequently (54 vs 40% of applications, p = 0.047) in patients treated with the standard catheter than in patients treated with the modified catheter. The use of catheter designed to increase the surface area of electrode-tissue contact allows more radiofrequency energy to be delivered before a rise in impedance occurs and appears to increase the effectiveness of radiofrequency ablation of the AV junction.
引用
收藏
页码:142 / 147
页数:6
相关论文
共 13 条
  • [1] THE EFFECT OF ELECTRODE DESIGN ON THE EFFICIENCY OF DELIVERY OF RADIOFREQUENCY ENERGY TO CARDIAC TISSUE INVITRO
    BLOUIN, LT
    MARCUS, FI
    [J]. PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1989, 12 (01): : 136 - 143
  • [2] THE PERCUTANEOUS CARDIAC MAPPING AND ABLATION REGISTRY - FINAL SUMMARY OF RESULTS
    EVANS, GT
    SCHEINMAN, MM
    SCHEINMAN, MM
    ZIPES, DP
    BENDITT, D
    BREITHARDT, G
    CAMM, AJ
    ELSHERIF, N
    FISHER, J
    FONTAINE, G
    LEVY, S
    PRYSTOWSKY, E
    JOSEPHSON, M
    MORADY, F
    RUSKIN, J
    [J]. PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1988, 11 (11): : 1621 - 1626
  • [3] TISSUE HEATING DURING RADIOFREQUENCY CATHETER ABLATION - A THERMODYNAMIC MODEL AND OBSERVATIONS IN ISOLATED PERFUSED AND SUPERFUSED CANINE RIGHT VENTRICULAR FREE WALL
    HAINES, DE
    WATSON, DD
    [J]. PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1989, 12 (06): : 962 - 976
  • [4] HOLT PM, 1988, CATHETER ABLATION CA, P1
  • [5] Hoyt RH, 1986, J APPL CARDIOL, V1, P469
  • [6] CLOSED CHEST CATHETER DESICCATION OF THE ATRIOVENTRICULAR JUNCTION USING RADIOFREQUENCY ENERGY - A NEW METHOD OF CATHETER ABLATION
    HUANG, SK
    BHARATI, S
    GRAHAM, AR
    LEV, M
    MARCUS, FI
    ODELL, RC
    [J]. JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1987, 9 (02) : 349 - 358
  • [7] RADIOFREQUENCY CATHETER ABLATION OF THE LEFT AND RIGHT VENTRICLES - ANATOMIC AND ELECTROPHYSIOLOGIC OBSERVATIONS
    HUANG, SKS
    GRAHAM, AR
    WHARTON, K
    [J]. PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1988, 11 (04): : 449 - 459
  • [8] CATHETER ABLATION OF THE ATRIOVENTRICULAR JUNCTION WITH RADIOFREQUENCY ENERGY
    LANGBERG, JJ
    CHIN, MC
    ROSENQVIST, M
    COCKRELL, J
    DULLET, N
    VANHARE, G
    GRIFFIN, JC
    SCHEINMAN, MM
    [J]. CIRCULATION, 1989, 80 (06) : 1527 - 1535
  • [9] LANGBERG JJ, IN PRESS PACE
  • [10] LAVERGNE T, 1986, LANCET, V2, P858