UTILITY OF TRANSTELEPHONIC ECG MONITORING IN EVALUATING DISCHARGES FROM A 3RD-GENERATION IMPLANTABLE CARDIOVERTER-DEFIBRILLATOR

被引:2
作者
HAMER, ME
WHARTON, JM
PAGE, RL
机构
[1] Duke University Medical Center, Durham, NC
关键词
ELECTROCARDIOGRAPHIC MONITOR; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR;
D O I
10.1016/0022-0736(93)90043-D
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
The authors report the case of a 60-year-old man with a third-generation implantable cardioverter defibrillator whose rate-sensing lead malfunction was diagnosed only after surface loop electrocardiographic (ECG) monitor recording. The stored R-R intervals suggested an irregular ventricular rhythm, but the loop ECG monitor demonstrated inappropriate shocks delivered during sinus rhythm.