AN IMPROVED X-RAY SPECTROMETRIC METHOD FOR THE DETERMINATION OF BROMIDE IN WHOLE-BLOOD OF WORKERS OCCUPATIONALLY EXPOSED TO METHYL-BROMIDE

被引:4
作者
HURST, JA
TONKS, CE
GEYER, R
机构
[1] Laboratory Services Unit, WorkCover Authority of NSW, Thomleigh, NSW, 2120
关键词
D O I
10.1093/jat/18.3.147
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
An X-ray spectrometric method using a sequential wavelength dispersive instrument is described for the analysis of bromide in whole blood from workers occupationally exposed to methyl bromide. The method involves the direct deposition of 100 μL of whole blood onto filter paper discs. Calibration of the instrument is achieved using the standard addition technique for bromide concentrations up to 50 mg/L. Precision studies at concentrations of 6.1, 16.4, and 26.9 mg/L gave relative standard deviation values of 5.8, 4.1, and 2.8%, respectively. The detection limit for the method is 1.2 mg/L. © 1994, Oxford University Press.
引用
收藏
页码:147 / 149
页数:3
相关论文
共 18 条
[1]   DETERMINATION OF IODINE AND BROMINE IN PLASMA AND URINE BY INDUCTIVELY COUPLED PLASMA MASS-SPECTROMETRY [J].
ALLAIN, P ;
MAURAS, Y ;
DOUGE, C ;
JAUNAULT, L ;
DELAPORTE, T ;
BEAUGRAND, C .
ANALYST, 1990, 115 (06) :813-815
[2]   GAS-CHROMATOGRAPHIC METHOD FOR DETERMINATION OF INCREASED BROMIDE CONCENTRATIONS IN BLOOD [J].
ARCHER, AW .
ANALYST, 1972, 97 (1155) :428-&
[3]  
BASELT RC, 1989, DISPOSITION TOXIC DR, P542
[4]   BROMIDE MEASUREMENT IN SERUM AND URINE BY AN IMPROVED GAS-CHROMATOGRAPHIC METHOD [J].
CORINA, DL ;
BALLARD, KE ;
GRICE, D ;
EADE, OE ;
LUCAS, K .
JOURNAL OF CHROMATOGRAPHY, 1979, 162 (03) :382-387
[5]  
DEJONG P, 1983, CLIN CHIM ACTA, V132, P63, DOI 10.1016/0009-8981(83)90233-4
[6]  
GOEWIE CE, 1985, J CHROMATOGR, V334, P157
[7]  
MATHIES JC, 1960, NORELCO REPTR, V7, P134
[8]  
MILLER ME, 1984, CLIN CHEM, V30, P781
[9]  
NATELSON S, 1962, CLIN CHEM, V8, P630
[10]  
PARMEGGIANI L, 1989, ENCY OCCUPATIONAL HL, V1, P330