TIME-SERIES METHODS IN THE MONITORING OF INTRACRANIAL-PRESSURE .1. PROBLEMS, SUGGESTIONS FOR A MONITORING SCHEME AND REVIEW OF APPROPRIATE TECHNIQUES

被引:13
作者
ALLEN, R [1 ]
机构
[1] UNIV LEEDS, DEPT ANAESTHESIA, LEEDS LS2 9JT, W YORKSHIRE, ENGLAND
来源
JOURNAL OF BIOMEDICAL ENGINEERING | 1983年 / 5卷 / 01期
关键词
SIGNAL FILTERING AND PREDICTION - Medical Applications - STATISTICAL METHODS - Time Series Analysis;
D O I
10.1016/0141-5425(83)90073-0
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Intracranial hypertension is a dangerous condition and poses a constant threat to patients suffering, for example, a severe head injury. Anticipation of this condition can, however, be difficult. This paper introduces problems that are created in patient management by the pathophysiology of intracranial hypertension and describes methods that have been devised to anticipate the condition, indicating their limitations. A scheme incorporating trend detection and prediction is then suggested for monitoring intracranial pressure (ICP). Several techniques from the fields of time series analysis and control engineering were selected as candidates for the basis of the scheme. These are described and their strengths and weaknesses discussed in the context of the ICP monitoring scheme. The application of time series methods to patient monitoring is reviewed in parallel with these descriptions to indicate the momentum of developments in this area of work. This theoretical background has enabled the direction of further work to be established. In particular it is suggested that adaptive methods are necessary to provide a satisfactory response to significant changes in the measured variable, to avoid repeated alarms and to allow useful short-term predictions of the variable to be made.
引用
收藏
页码:5 / 17
页数:13
相关论文
共 77 条
[1]   MICROCOMPUTER ASSISTANCE IN CLINICAL MONITORING OF INTRA-CRANICAL PRESSURE [J].
ALLEN, R .
MEDICAL & BIOLOGICAL ENGINEERING & COMPUTING, 1981, 19 (03) :349-355
[2]   A DISTRIBUTED COMPUTER NETWORK USING MICROCOMPUTERS FOR ONLINE DATA-PROCESSING IN CLINICAL MONITORING AND MEDICAL-RESEARCH [J].
ALLEN, R .
JOURNAL OF BIOMEDICAL ENGINEERING, 1980, 2 (04) :299-304
[3]  
ALLEN R, 1978, THESIS U LEEDS
[4]  
Anderson B., 1979, OPTIMAL FILTERING
[5]   SYSTEM IDENTIFICATION - SURVEY [J].
ASTROM, KJ ;
EYKHOFF, P .
AUTOMATICA, 1971, 7 (02) :123-+
[6]  
AVEZAAT CJJ, 1976, INTRACRANIAL PRESSUR, V3, P308
[7]  
Becker DP, 1972, INTRACRANIAL PRESSUR, P309
[8]  
Beneken J E, 1979, J Biomed Eng, V1, P185, DOI 10.1016/0141-5425(79)90041-4
[9]  
Box GE., 1970, SAN FRANCISCO HOLDAN
[10]  
BUSHMAN JA, 1976, REAL TIME COMPUTING, P167