DISPLAY OF CT STUDIES ON A 2-SCREEN ELECTRONIC WORKSTATION VERSUS A FILM PANEL ALTERNATOR - SENSITIVITY AND EFFICIENCY AMONG RADIOLOGISTS

被引:28
作者
FOLEY, WD
JACOBSON, DR
TAYLOR, AJ
GOODMAN, LR
STEWART, ET
GURNEY, JW
STROKA, D
机构
[1] MED COLL WISCONSIN,MANAGEMENT ENGN SECT,MILWAUKEE,WI 53226
[2] UNIV NEBRASKA,MED CTR,DEPT RADIOL,OMAHA,NE 68105
关键词
Computed tomography (CT); image display and recording; diagnostic radiology; observer performance; images; display; interpretation;
D O I
10.1148/radiology.174.3.2305060
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Prototype electronic workstations incorporated in networks linking computed tomographic (CT) and magnetic resonance imaging systems are being developed. The authors compared observer efficiency and sensitivity in reading body CT studies from a two-screen workstation (1,000-line monitors and 12-bit dynamic range in image memory) and conventional film panel alternator. The two-screen workstation displayed 32 images at a matrix resolution of 256 x 256 or eight images at a matrix resolution of 512 x 512 simultaneously. Ninety-six images with a matrix resolution of 512 x 512 could be displayed simultaneously at the film panel alternator. Four observers read images from 20 cases, 10 with repeat examiniations, in a randomized viewing sequence. There was an average of 32 images per case. Reporting time was less with the film panel alternator (average, 5.08 minutes) than with the workstation (average, 6.66 minutes). There was improved sensitivity for all observers in reading from the film panel alternator (range, 1%-12%) (P < .05). In complex cases evaluated by means of body CT, the current prototype two-screen electronic workstation is limited by display capabilities.
引用
收藏
页码:769 / 773
页数:5
相关论文
共 14 条
[1]  
FOLEY WD, 1986, INVEST RADIOL, V21, P305
[2]   THE EFFECT OF VARYING SPATIAL-RESOLUTION ON THE DETECTABILITY OF DIFFUSE PULMONARY NODULES - ASSESSMENT WITH DIGITIZED CONVENTIONAL RADIOGRAPHS [J].
FOLEY, WD ;
WILSON, CR ;
KEYES, GS ;
DIBIANCA, FA ;
SCANLON, GT ;
SCHLEUTER, D ;
LAWSON, TL .
RADIOLOGY, 1981, 141 (01) :25-31
[3]  
FOLEY WD, 1988, 1987 P CHEST IM C MA, P96
[4]  
FRASER RG, 1983, RADIOLOGY, V148, P1
[5]   PNEUMOTHORAX AND OTHER LUNG-DISEASES - EFFECT OF ALTERED RESOLUTION AND EDGE ENHANCEMENT ON DIAGNOSIS WITH DIGITIZED RADIOGRAPHS [J].
GOODMAN, LR ;
FOLEY, WD ;
WILSON, CR ;
TIKOFSKY, RS ;
HOFFMANN, RG .
RADIOLOGY, 1988, 167 (01) :83-88
[6]   DIGITAL AND CONVENTIONAL CHEST IMAGES - OBSERVER PERFORMANCE WITH FILM DIGITAL RADIOGRAPHY SYSTEM [J].
GOODMAN, LR ;
FOLEY, WD ;
WILSON, CR ;
RIMM, AA ;
LAWSON, TL .
RADIOLOGY, 1986, 158 (01) :27-33
[8]   DIGITAL PROJECTION RADIOGRAPHY OF THE CHEST [J].
KUNDEL, HL .
RADIOLOGY, 1986, 158 (01) :274-276
[9]  
KUNDEL HL, 1986, RADIOL CLIN N AM, V24, P69
[10]   LOW-DOSE FLYING SPOT DIGITAL RADIOGRAPHY OF THE CHEST - SENSITIVITY STUDIES [J].
KUSHNER, DC ;
CLEVELAND, RH ;
HERMAN, TE ;
MCLOUD, TC ;
WALTMAN, AC ;
SHEPARD, JAO ;
DEDRICK, CG ;
KOPANS, DB ;
GREENE, RE .
RADIOLOGY, 1987, 163 (03) :685-688