CORONARY ANGIOGRAPHY WITH SYNCHROTRON-RADIATION

被引:4
作者
DIX, WR
BESCH, HJ
GRAEFF, W
HAMM, CW
ILLING, G
KUPPER, W
LOHMANN, M
MEINERTZ, T
MENK, RH
REIME, B
RUST, C
SCHILDWACHTER, L
TAFELMEIER, U
WALENTA, AH
机构
[1] UNIV GESAMTHSCH SIEGEN,FACHBEREICH PHYS 7,D-57068 SIEGEN,GERMANY
[2] UNIV HAMBURG,KRANKENHAUS EPPENDORF,MED KLIN,KARDIOL ABT,D-20251 HAMBURG,GERMANY
[3] HERZ KREISLAUF KLIN,D-29549 BAD BEVENSON,GERMANY
来源
JOURNAL DE PHYSIQUE IV | 1994年 / 4卷 / C9期
关键词
D O I
10.1051/jp4:1994949
中图分类号
O4 [物理学];
学科分类号
0702 ;
摘要
Worldwide several systems for Digital Subtraction Angiography in energy subtraction mode (dichromography) are developed, One of the systems which allows investigations of patients - the system NIKOS in the Hamburger Synchrotronstrahlungslabor HASYLAB at DESY is described. The system consists out of sis main parts: The 20 pole wiggler HXRWI in the beamline W2 of the storage ring DORIS, the two beam monochromator for the two 12.5 cm wide monochromatic X-ray beams, the safety system, the fast scanning device, the fast low-noise two-line ionization chamber and the computer system. Since 1990 patients were investigated with the system NIKOS, him of the work is to visualize coronary arteries down to 1 mm diameter with an iodine mass,density of 1 mg/cm(2), thus allowing non-invasive investigations by intravenous application of the contrast medium. The two images for subtraction are simultaneously taken with photon energies just below and above the iodine K-edge (33.17 keV) in a line scan mode. Intravenous angiograms from investigations with version III of the system NIKOS are presented and the next steps for getting an increased image quality are described.
引用
收藏
页码:279 / 286
页数:8
相关论文
共 13 条
[1]  
BESCH HJ, 1993, PHYS MEDICA, V9, P171
[2]   CORONARY ANGIOGRAPHY USING SYNCHROTRON RADIATION - STUDIES IN HUMAN-SUBJECTS WITH THE SYSTEM NIKOS-II [J].
DIX, WR ;
ENGELKE, K ;
GRAEFF, W ;
HAMM, C ;
HEUER, J ;
KAEMPF, B ;
KUPPER, W ;
LOHMANN, M ;
REIME, B ;
REUMANN, R .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1992, 314 (02) :307-315
[3]  
DOLBNYA IP, 1990, PHYS MEDICA, V6, P313
[4]  
ELLEAUME H, 1994, FEB WORKSH ESRF GREN
[5]   HARWI - A HARD X-RAY WIGGLER BEAM AT DORIS [J].
GRAEFF, W ;
BITTNER, L ;
BREFELD, W ;
HAHN, U ;
HEINTZE, G ;
HEUER, J ;
KOUPTSIDIS, J ;
PFLUGER, J ;
SCHWARTZ, M ;
WEINER, EW ;
WROBLEWSKI, T .
REVIEW OF SCIENTIFIC INSTRUMENTS, 1989, 60 (07) :1457-1459
[6]  
GRAEFF W, 1991, HDB SYNCHROTRON RAD, V4, P407
[7]  
HYODO K, 1991, HDB SYNCHROTRON RAD, V4, P55
[8]  
KUPPER W, 1988, SYNCHROTRON RAD APPL, P165
[9]   SYNCHROTRON RADIATION CORONARY ANGIOGRAPHY WITH A DUAL-BEAM, DUAL-DETECTOR IMAGING-SYSTEM [J].
RUBENSTEIN, E ;
GIACOMINI, JC ;
GORDON, HJ ;
THOMPSON, AC ;
BROWN, G ;
HOFSTADTER, R ;
THOMLINSON, W ;
ZEMAN, HD .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1990, 291 (1-2) :80-85
[10]  
RUST C, THESIS U HAMBRUG FRG