A NEW APPROACH TOWARDS DEFIBRILLATION ELECTRODES - HIGHLY CONDUCTIVE ISOTROPIC CARBON-FIBERS

被引:8
作者
ALT, E [1 ]
THERES, H [1 ]
HEINZ, M [1 ]
ALBRECHT, K [1 ]
GEORG, H [1 ]
BLOEMER, H [1 ]
机构
[1] TECH UNIV MUNICH,KLINIKUM RECHTS ISAR,DEPT MED ELECTR,W-8000 MUNICH 80,GERMANY
来源
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY | 1991年 / 14卷 / 11期
关键词
DEFIBRILLATION; ELECTRODES; CARBON FIBERS;
D O I
10.1111/j.1540-8159.1991.tb02791.x
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
A new carbon fiber material was studied for its potential benefit in defibrillation electrodes. Miniaturization of the fiber production can result in small strands with extremely large surface areas. Two carbon fiber patches and a single carbon fiber bundle were used for the purposes of this study, and the results were compared to those obtained with conventional titanium/mesh patch electrodes. Tests performed in a saline filled tank revealed considerably lower resistances through the carbon material when compared to standard mesh electrodes. There was also a higher peak current flow with lower voltage. The carbon fibers exhibited lower impedance for defibrillation and this may be related to low polarization known to occur with carbon materials. This new carbon material may prove to be useful and further research is required.
引用
收藏
页码:1923 / 1928
页数:6
相关论文
共 6 条
[1]  
FISHER JD, 1991, NEW PERSPECTIVES CAR, V2, P359
[2]  
IDEKER RE, 1987, NONPHARMACOLOGICAL T, P449
[3]   STANDBY AUTOMATIC DEFIBRILLATOR - AN APPROACH TO PREVENTION OF SUDDEN CORONARY DEATH [J].
MIROWSKI, M ;
MOWER, MM ;
STAEWEN, WS ;
TABATZNIK, B ;
MENDELOFF, AI .
ARCHIVES OF INTERNAL MEDICINE, 1970, 126 (01) :158-+
[4]   TERMINATION OF MALIGNANT VENTRICULAR ARRHYTHMIAS WITH AN IMPLANTED AUTOMATIC DEFIBRILLATOR IN HUMAN-BEINGS [J].
MIROWSKI, M ;
REID, PR ;
MOWER, MM ;
WATKINS, L ;
GOTT, VL ;
SCHAUBLE, JF ;
LANGER, A ;
HEILMAN, MS ;
KOLENIK, SA ;
FISCHELL, RE ;
WEISFELDT, ML .
NEW ENGLAND JOURNAL OF MEDICINE, 1980, 303 (06) :322-324
[5]  
MOND HG, 1991, NEW PERSPECTIVES CAR, P133
[6]   ELECTROPHYSIOLOGICAL EFFECTS OF MONOPHASIC AND BIPHASIC STIMULI IN NORMAL AND INFARCTED DOGS [J].
WHARTON, JM ;
RICHARD, VJ ;
MURRY, CE ;
DIXON, EG ;
REIMER, KA ;
MEADOR, J ;
SMITH, WM ;
IDEKER, RE .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1990, 13 (09) :1158-1172