THE EVALUATION OF A NEW INTRAVASCULAR BLOOD-GAS MONITORING-SYSTEM IN THE PIG

被引:34
作者
CLUTTONBROCK, TH [1 ]
FINK, S [1 ]
MARKLE, D [1 ]
LUTHRA, AJ [1 ]
HENDRY, SP [1 ]
机构
[1] PFIZER HOSP,BIOMED SENSORS LTD,PROD GRP,HIGH WYCOMBE,ENGLAND
来源
JOURNAL OF CLINICAL MONITORING | 1994年 / 10卷 / 06期
关键词
EQUIPMENT; INTRAVASCULAR SENSORS; MONITORING; BLOOD GAS ANALYSIS;
D O I
10.1007/BF01618420
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
Objective. Our objective was to investigate the accuracy of a new intravascular blood gas sensor, the Paratrend 7 (P7) (Biomedical Sensors Ltd, Pfizer Hospital Products Group, High Wycombe, England) in a porcine model. Methods. A total of 12 sensors were inserted into 10 animals under total intravenous anesthesia. Changes in blood gas chemistry were produced over a wide range by manipulagting the inspired oxygen and carbon dioxide concentrations and by adjustments in minute ventilation. Blood gas samples (BGA) were taken and analyzed during periods of stability; the results obtained were compared with the readings from the intravascular sensor. Results. A total of 292 blood gas samples were taken and analyzed for pHa, Paco2, and Po2; the results were compared with the readings from the intravascular sensor. Correlation coefficients of r = 0.98 for Pco2 and r = 0.99 for Po2 were obtained. Analysis of bias and precision as mean +/- SD of the difference (P7 - BGA) gave the following results: pH bias = -0.03, precision = +/- 0.04; Pco2 bias = 0.65 mm Hg, precision = +/- 3.1 mm Hg; and Po2 bias = -6.50 mm Hg, precision = +/- 0.6 mm Hg. No problems with clot formation on the sensor were seen, and the sensors did not appear to show the ''wall effect'' seen with other systems. Conclusions. The results obtained were well within the requirements for a clinically useful blood gas monitoring system.
引用
收藏
页码:387 / 391
页数:5
相关论文
共 8 条
  • [1] STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT
    BLAND, JM
    ALTMAN, DG
    [J]. LANCET, 1986, 1 (8476) : 307 - 310
  • [2] EVALUATION OF A DUAL-FUNCTION PH AND PCO-2 INVIVO SENSOR
    COON, RL
    LAI, NCJ
    KAMPINE, JP
    [J]. JOURNAL OF APPLIED PHYSIOLOGY, 1976, 40 (04) : 625 - 629
  • [3] ELSER R, 1990, IFCC AACC, P343
  • [4] ASSESSING PRECISION AND ACCURACY IN BLOOD-GAS PROFICIENCY TESTING
    HANSEN, JE
    CASABURI, R
    CRAPO, RO
    JENSEN, RL
    [J]. AMERICAN REVIEW OF RESPIRATORY DISEASE, 1990, 141 (05): : 1190 - 1193
  • [5] PROGRESS IN THE DEVELOPMENT OF A FLUORESCENT INTRAVASCULAR BLOOD-GAS SYSTEM IN MAN
    MAHUTTE, CK
    SASSOON, CSH
    MURO, JR
    HANSMANN, DR
    MAXWELL, TP
    MILLER, WW
    YAFUSO, M
    [J]. JOURNAL OF CLINICAL MONITORING, 1990, 6 (02): : 147 - 157
  • [6] THE PERFORMANCE-CHARACTERISTICS OF AN INTRA-ARTERIAL OXYGEN-ELECTRODE
    RITHALIA, SVS
    BENNETT, PJ
    TINKER, J
    [J]. INTENSIVE CARE MEDICINE, 1981, 7 (06) : 305 - 307
  • [7] PRELIMINARY EVALUATION OF AN INTRA-ARTERIAL BLOOD-GAS SYSTEM IN DOGS AND HUMANS
    SHAPIRO, BA
    CANE, RD
    CHOMKA, CM
    BANDALA, LE
    PERUZZI, WT
    [J]. CRITICAL CARE MEDICINE, 1989, 17 (05) : 455 - 460
  • [8] 1992, FED REGISTER, V40, P70002