ANALYSIS OF PROTEIN-FRACTIONS BY MICROPREPARATIVE CAPILLARY ISOELECTRIC-FOCUSING AND MATRIX-ASSISTED LASER-DESORPTION TIME-OF-FLIGHT MASS-SPECTROMETRY

被引:58
作者
FORET, F [1 ]
MULLER, O [1 ]
THORNE, J [1 ]
GOTZINGER, W [1 ]
KARGER, BL [1 ]
机构
[1] NORTHEASTERN UNIV,BARNETT INST,BOSTON,MA 02115
关键词
D O I
10.1016/0021-9673(95)00621-S
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
In this study, the use of capillary isoelectric focusing (cIEF) as a micropreparative tool for protein analysis by matrix-assisted laser desorption time-of-flight mass spectrometry (MALDI-TOF-MS) is demonstrated. A newly designed, automated, collection interface equipped with a fiber-optic UV detector and a sheath flow connection was employed for collection of protein fractions. Multiple fractions were collected during a single cIEF run and further analyzed by MALDI-TOF-MS for mass assignment. The feasibility of the method was tested with a mixture of model proteins with different isoelectric points and molecular masses, and with variants of human hemoglobins differing in pI, but with negligible difference in M(r). Some practical considerations of the collection procedure and subsequent TOF analysis are presented.
引用
收藏
页码:157 / 166
页数:10
相关论文
共 19 条
[1]  
Bunn H. F., 1986, HEMOGLOBIN MOL GENET, P61
[2]   ISOELECTRIC-FOCUSING BY FREE SOLUTION CAPILLARY ELECTROPHORESIS [J].
CHEN, SM ;
WIKTOROWICZ, JE .
ANALYTICAL BIOCHEMISTRY, 1992, 206 (01) :84-90
[3]  
GOTZINGER W, 1995, 7TH INT S HIGH PERF, P637
[4]  
GRIMM R, 1995, 7TH INT S HIGH PERF, P633
[5]  
Hames B.D., 1990, GEL ELECTROPHORESIS
[6]   ADAPTATION OF THE EQUIPMENT FOR HIGH-PERFORMANCE ELECTROPHORESIS TO ISOELECTRIC-FOCUSING [J].
HJERTEN, S ;
ZHU, MD .
JOURNAL OF CHROMATOGRAPHY, 1985, 346 (OCT) :265-270
[7]   FAST AND HIGH-RESOLUTION ANALYSIS OF HUMAN-SERUM TRANSFERRIN BY HIGH-PERFORMANCE ISOELECTRIC-FOCUSING IN CAPILLARIES [J].
KILAR, F ;
HJERTEN, S .
ELECTROPHORESIS, 1989, 10 (01) :23-29
[8]  
KYIOHITO S, 1994, ANAL CHEM, V66, P9
[9]   SIMPLE APPROACH TO ELIMINATING DISTURBANCES IN ISOELECTRIC-FOCUSING CAUSED BY THE PRESENCE OF SALTS [J].
LIAO, JL ;
ZHANG, R .
JOURNAL OF CHROMATOGRAPHY A, 1994, 684 (01) :143-148
[10]   HIGH-RESOLUTION TANDEM MASS-SPECTROMETRY OF LARGE BIOMOLECULES [J].
LOO, JA ;
QUINN, JP ;
RYU, SI ;
HENRY, KD ;
SENKO, MW ;
MCLAFFERTY, FW .
PROCEEDINGS OF THE NATIONAL ACADEMY OF SCIENCES OF THE UNITED STATES OF AMERICA, 1992, 89 (01) :286-289