SCANNING ELECTRON-MICROSCOPY OF SURFACE IRREGULARITIES AND THROMBOGENESIS OF POLYURETHANE AND POLYETHYLENE CORONARY CATHETERS

被引:59
作者
BOURASSA, MG
CANTIN, M
SANDBORN, EB
PEDERSON, E
机构
[1] MONTREAL HEART INST, MONTREAL H1T 1C8, QUEBEC, CANADA
[2] UNIV MONTREAL, SCH MED, DEPT PATHOL, MONTREAL 101, QUEBEC, CANADA
[3] UNIV MONTREAL, SCH MED, DEPT ANAT, MONTREAL 101, QUEBEC, CANADA
关键词
D O I
10.1161/01.CIR.53.6.992
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Following routine coronary arteriography, surface irregularities and thrombogenesis of the inner and outer wall of 6 Ducor polyurethane and 6 RPX polyethylene coronary catheters were studied by scanning EM. Polyurethane catheters had rough and highly irregular external and internal surfaces. All catheters showed adherent thrombi on their external surface. The internal surface of 3 catheters showed evidence of thrombosis. Polyethylene differed from polyurethane in several respects. Although the external surface had an irregular and wavelike appearance, the internal surface was smooth and regular. Polyethylene catheters (2) showed thrombi on their external surface. The internal surface of 1 catheter showed single platelets in 1 area. Internal and external surface irregularities play a major role in the initiation of thrombosis in and on intravascular catheters. They stress the need for high quality catheter materials with smooth and regular surfaces in the prevention of thromboembolic complications from coronary arteriography.
引用
收藏
页码:992 / 996
页数:5
相关论文
共 33 条
[1]   COMPLICATIONS OF CORONARY ARTERIOGRAPHY [J].
ADAMS, DF ;
FRASER, DB ;
ABRAMS, HL .
CIRCULATION, 1973, 48 (03) :609-618
[2]   COMPLICATIONS OF PERCUTANEOUS FEMORAL ARTERIAL CATHETERIZATION - PROSPECTIVE EVALUATION WITH DOPPLER ULTRASONIC VELOCITY DETECTOR [J].
BARNES, RW ;
PETERSEN, JL ;
KRUGMIRE, RB ;
STRANDNESS, DE .
AMERICAN JOURNAL OF CARDIOLOGY, 1974, 33 (02) :259-263
[3]  
BOURASSA MG, 1970, CAN MED ASSOC J, V102, P170
[4]   COMPLICATION RATE OF CORONARY ARTERIOGRAPHY - REVIEW OF 5250 CASES STUDIED BY A PERCUTANEOUS FEMORAL TECHNIQUE [J].
BOURASSA, MG ;
NOBLE, J .
CIRCULATION, 1976, 53 (01) :106-114
[5]   SELECTIVE CORONARY ARTERIOGRAPHY BY PERCUTANEOUS FEMORAL ARTERY APPROACH [J].
BOURASSA, MG ;
LESPERANCE, J ;
CAMPEAU, L .
AMERICAN JOURNAL OF ROENTGENOLOGY RADIUM THERAPY AND NUCLEAR MEDICINE, 1969, 107 (02) :377-+
[6]   REDUCTION OF SURGICAL COMPLICATION RATE BY USE OF A HYPOTHROMBOGENIC CATHETER COATING [J].
CRAMER, R ;
MOORE, R ;
AMPLATZ, K .
RADIOLOGY, 1973, 109 (03) :585-588
[7]   EMBOLIC CORONARY-ARTERY OCCLUSION IN PERCUTANEOUS TRANSFEMORAL CORONARY ARTERIOGRAPHY [J].
DELATORRE, A ;
JACOBS, D ;
ALEMAN, J ;
ANDERSON, GA .
AMERICAN HEART JOURNAL, 1973, 86 (04) :467-473
[9]   ARTERIAL THROMBUS FORMATION DURING CLINICAL PERCUTANEOUS CATHETERIZATION [J].
FORMANEK, G ;
FRECH, RS ;
AMPLATZ, K .
CIRCULATION, 1970, 41 (05) :833-&
[10]  
HAULT G, 1968, SURGERY, V63, P594