EVALUATION OF A COMPLETE HPLC SOLVENT OPTIMIZATION SYSTEM INVOLVING PIECE-WISE QUADRATIC MODELING

被引:19
作者
NAISHCHAMBERLAIN, PJ [1 ]
LYNCH, RJ [1 ]
机构
[1] PHILIPS SCI,YORK ST,CAMBRIDGE,ENGLAND
关键词
Column liquid chromatography; Piecewise quadratic modelling; Solven optimization;
D O I
10.1007/BF02261144
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
Interpretive methods are accepted to give the best possible results for selectivity optimization in HPLC. However the methods are very complex, and most work so far has been detailed academic studies. This paper describes an evaluation of a complete integrated system incorporating peak labelling, modelling of retention behaviour and calculation of response surfaces, with particular emphasis on the retention modelling. The peak labelling section has been discussed previously. A piece-wise quadratic function is investigated for the modelling of retention times across an isoeluotropic plane to effect selectivity optimization in HPLC. This requires 10 data-points on the isoeluotropic plane. The predicted global optimum and local optima are evaluated by comparison of calculated and experimental retention data, for a nine component sample. Seven interstitial points, distributed across the whole plane between the data-points, are similarly evaluated for a related sample. The typical error (in retention time) is less than 2%, often 1%, and the maximum error is 4.2%. At the global optimum the error was found be less than 1.3% for all 9 peaks. © 1990 Friedr. Vieweg & Sohn Verlagsgesellschaft mbH.
引用
收藏
页码:79 / 89
页数:11
相关论文
共 9 条
[1]  
BLAFFERT T, 1987, SCI COMPUTING AUTOMA
[2]  
GLASCH JL, 1980, J CHROMATOGR, V199, P57
[3]   A STUDY OF THE PRACTICAL LIMITATIONS OF PRINCIPAL COMPONENTS-ANALYSIS AND THE LABELING OF UNRESOLVED HPLC PEAKS [J].
NAISH, PJ ;
LYNCH, RJ ;
BLAFFERT, T .
CHROMATOGRAPHIA, 1989, 27 (7-8) :343-358
[4]   EFFECT OF MODEL INACCURACY ON SELECTIVITY OPTIMIZATION PROCEDURES IN REVERSED-PHASE LIQUID-CHROMATOGRAPHY [J].
SCHOENMAKERS, PJ ;
BLAFFERT, T .
JOURNAL OF CHROMATOGRAPHY, 1987, 384 :117-133
[5]   CRITERIA FOR COMPARING THE QUALITY OF CHROMATOGRAMS WITH GREAT VARIATIONS IN CAPACITY FACTORS [J].
SCHOENMAKERS, PJ .
JOURNAL OF LIQUID CHROMATOGRAPHY, 1987, 10 (8-9) :1865-1886
[6]  
SCHOENMAKERS PJ, 1986, J CHROMATOGRAPHY LIB, V35, P170
[7]  
SCHOENMAKERS PJ, 1986, J CHROMATOGRAPHY LIB, V35, P67
[8]  
SCHOENMAKERS PJ, 1986, J CHROMATOGRAPHY LIB, V35, P12
[9]  
SNYDER LR, 1979, INTRO MODERN LIQUID, P264