UNUSUAL ANALYTE ADSORPTION EFFECTS ON INERT LC COMPONENTS

被引:10
作者
HAMBLETON, P [1 ]
LOUGH, WJ [1 ]
MALTAS, J [1 ]
MILLS, MJ [1 ]
机构
[1] GLAXO RES & DEV LTD,DEPT DRUG METAB,GREENFORD UB6 0HE,MIDDX,ENGLAND
来源
JOURNAL OF LIQUID CHROMATOGRAPHY | 1995年 / 18卷 / 16期
关键词
D O I
10.1080/10826079508010445
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
The use of injector loops made of the inert polymer PEEK(TM) was studied for the analysis of non-steroidal anti-inflammatory drugs (NSAID) using microbore LC with peak compression. Surprisingly it was found that the use of PEEK not only failed to prevent adsorption of analyte in the loop but, in this instance, actually exacerbated the problem. The effect of loop overfill volume, competing NSAID in the sample solution and percentage organic component in the mobile phase on this phenomenon was investigated. Comparison was also made between PEEK(TM) and other materials used for injector loops.
引用
收藏
页码:3205 / 3217
页数:13
相关论文
共 9 条
[1]   A COMPARATIVE-STUDY OF LARGE VOLUME INJECTION TECHNIQUES FOR MICROBORE COLUMNS IN HPLC [J].
CLAESSENS, HA ;
KUYKEN, MAJ .
CHROMATOGRAPHIA, 1987, 23 (05) :331-336
[2]  
DOLAN JW, 1991, LC GC EUR, V4, P16
[3]  
HOFFMAN NE, 1991, J LIQ CHROMATOGR, V14, P65
[4]  
INMAN EL, 1989, J CHROMATOGR, V465, P236
[5]  
LOUGH WJ, UNPUB ANAL ABSORPTIO
[6]   COMMENTS ON SOLUTE-SOLVENT INTERACTIONS IN QUANTITATIVE HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY [J].
MACLEOD, SK ;
FAGAN, DT ;
SCHOLL, JP .
JOURNAL OF CHROMATOGRAPHY, 1990, 502 (01) :236-240
[7]  
OTERO GCF, 1993, J CHROMATOGR A, V654, P87
[8]   TRACE ANALYSIS BY PEAK COMPRESSION SAMPLING OF A LARGE SAMPLE VOLUME ON MICROBORE COLUMNS IN LIQUID-CHROMATOGRAPHY [J].
SLAIS, K ;
KOURILOVA, D ;
KREJCI, M .
JOURNAL OF CHROMATOGRAPHY, 1983, 282 (DEC) :363-370
[9]  
SPRUCE B, 1992, TROUBLE SHOOTING GUI, P12