NEW AUTOMATED HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHIC ANALYSIS OF CYCLOSPORINE-A AND CYCLOSPORINE-G IN HUMAN-SERUM

被引:28
作者
WALLEMACQ, PE
LESNE, M
机构
来源
JOURNAL OF CHROMATOGRAPHY-BIOMEDICAL APPLICATIONS | 1987年 / 413卷
关键词
D O I
10.1016/0378-4347(87)80220-7
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
An automated isocratic high-performance liquid chromatographic (HPLC) method is described for the determination of cyclosporin A and G in human serum. This method involves the use of an automated solid-liquid extraction procedure following rapid protein precipitation with acetonitrile. The use of a disposable C8 extraction cartridge allows a good recovery of cyclosporine (87%) from serum and a detection limit of 20 ng/ml with good reproducibility using 0.5 ml of sample. This method can also be adapted to whole blood measurements. The choice of a 3-.mu.m cyano analytical column and of the mobile phase hexane-isopropanol (85:15) permitted a low column temperature (50.degree.C), a low flow-rate (0.6 ml/min) and a short run time (14 min). This method allows the accurate and fast routine monitoring of cyclosporine by HPLC, which is particularly important in hepatic transplantations.
引用
收藏
页码:131 / 140
页数:10
相关论文
共 28 条
[1]   MEASUREMENT OF CYCLOSPORIN BY HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY FOLLOWING CHARCOAL ADSORPTION FROM WHOLE-BLOOD [J].
ARAVIND, MK ;
MICELI, JN ;
KAUFFMAN, RE .
JOURNAL OF CHROMATOGRAPHY, 1985, 344 (NOV) :428-432
[2]  
BOREL JF, 1976, IMMUNOLOGY, V31, P631
[3]  
BURCKART GJ, 1985, AM ASS CLIN CHEM, V6, P1
[4]  
CARRUTHERS SG, 1983, CLIN CHEM, V29, P180
[5]   SERUM CYCLOSPORIN LEVELS, HEPATIC DRUG-METABOLISM AND RENAL TUBULOTOXICITY [J].
CUNNINGHAM, C ;
GAVIN, MP ;
WHITING, PH ;
BURKE, MD ;
MACINTYRE, F ;
THOMSON, AW ;
SIMPSON, JG .
BIOCHEMICAL PHARMACOLOGY, 1984, 33 (18) :2857-2861
[6]  
Donatsch P, 1981, J Immunoassay, V2, P19, DOI 10.1080/01971528108062989
[7]   SIMPLIFIED LIQUID-CHROMATOGRAPHIC ANALYSIS FOR CYCLOSPORIN-A IN BLOOD AND PLASMA WITH USE OF RAPID EXTRACTION [J].
GARRAFFO, R ;
LAPALUS, P .
JOURNAL OF CHROMATOGRAPHY, 1985, 337 (02) :416-422
[8]   MODIFIED COLUMN-SWITCHING HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHIC METHOD FOR THE MEASUREMENT OF CYCLOSPORINE IN SERUM [J].
GMUR, DJ ;
YEE, GC ;
KENNEDY, MS .
JOURNAL OF CHROMATOGRAPHY, 1985, 344 (NOV) :422-427
[9]   SEMI-AUTOMATED HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHIC DETERMINATION OF CYCLOSPORINE A IN WHOLE-BLOOD USING ONE-STEP SAMPLE PURIFICATION AND COLUMN-SWITCHING [J].
HAMILTON, G ;
ROTH, E ;
WALLISCH, E ;
TICHY, F .
JOURNAL OF CHROMATOGRAPHY, 1985, 341 (02) :411-419
[10]  
HIESTAND PC, 1985, IMMUNOLOGY, V55, P249