DEVELOPMENT OF AN ISFET BASED HEPARIN SENSOR USING THE ION-STEP MEASURING METHOD

被引:16
作者
VANKERKHOF, JC [1 ]
BERGVELD, P [1 ]
SCHASFOORT, RBM [1 ]
机构
[1] TNO,3700 AJ ZEIST,NETHERLANDS
关键词
BIOSENSOR; ISFET; HEPARIN; PROTAMINE;
D O I
10.1016/0956-5663(93)80031-J
中图分类号
Q6 [生物物理学];
学科分类号
071011 ;
摘要
The ion-step measuring method was used to determine heparin concentrations in PBS and blood plasma. Heparin is a sulphated polysaccharide which is clinically used as a drug to prevent the clotting of blood. The measuring method is based on detection of changes in charge density in a porous membrane which is deposited on the gate of an ISFET. Protamine was used as affinity ligand in the membrane. In PBS a linear relation was found between the heparin concentration and the ISFET response. The incubation time was reduced from 18 h to 15 min by increasing the porosity of the membrane. The results of the measurements in blood plasma show a significant nonspecific binding of plasma components in the membrane. Suggestions are given to prevent this nonspecific adsorption. The results described in this paper show a detection limit for the ion-step measuring method of at least 5 +/- 10(-11) Mol/l which is promising for future practical applications.
引用
收藏
页码:463 / 472
页数:10
相关论文
共 14 条
[1]  
ABILDGAARD U, 1989, HEPARIN, pCH23
[2]  
BARROWCLIFFE TW, 1989, HEPARIN, pCH18
[3]   THE OPERATION OF AN ISFET AS AN ELECTRONIC DEVICE [J].
BERGVELD, P .
SENSORS AND ACTUATORS, 1981, 1 (01) :17-29
[4]  
BJORK L, 1989, HEPARIN, pCH11
[5]  
CASU B, 1989, HEPARIN, pCH2
[6]  
CUNDALL RB, 1989, MAKROMOL CHEM, V183, P849
[7]  
EIJKEL JCT, 1992, EUROSENSORS, V6, P336
[8]  
HIRSH J, 1991, NEW ENGL J MED, V324, P1565
[9]  
JAQUES LB, 1990, J LAB CLIN MED, V115, P422
[10]   APPEARANCE OF HEPARIN ANTITHROMBIN-ACTIVE CHAINS INVIVO AFTER INJECTION OF COMMERCIAL HEPARIN AND IN ANAPHYLAXIS [J].
LEVY, SW ;
JAQUES, LB .
THROMBOSIS RESEARCH, 1978, 13 (03) :429-441