DIRECT DETERMINATION OF MAGNESIUM AT THE 1-PERCENT LEVEL IN SOLID SAMPLES BY GRAPHITE-FURNACE ATOMIC-ABSORPTION SPECTROMETRY

被引:10
作者
BELARRA, MA
LAVILLA, I
CASTILLO, JR
机构
[1] Department of Analytical Chemistry, University of Zaragoza
关键词
POLY(VINYL CHLORIDE) ANALYSIS; DIRECT ANALYSIS; SOLID SAMPLE; MAGNESIUM DETERMINATION; GRAPHITE FURNACE ATOMIC ABSORPTION SPECTROMETRY;
D O I
10.2116/analsci.11.651
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Magnesium can be determined at levels above 1% in poly (vinyl chloride) (PVC) by direct analysis of the solid sample using graphite furnace atomic absorption spectrometry (GFAAS) and working at the 383.4 nm line. The analytical conditions used permit the determination of magnesium at a concentration of 2% by interpolation on a calibration graph run for the aqueous solution. No special sample introduction system, platform or matrix modification is required. The relative standard deviation thus achieved is ca. 7%.
引用
收藏
页码:651 / 656
页数:6
相关论文
共 27 条
[1]   DIRECT DETERMINATION OF COPPER AT TRACE LEVELS IN SOLID SAMPLES OF ANIMAL FEED USING ELECTROTHERMAL ATOMIC-ABSORPTION SPECTROMETRY [J].
ANZANO, JM ;
MARTINEZGARBAYO, MP ;
BELARRA, MA ;
CASTILLO, JR .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1994, 9 (02) :125-128
[2]   ON THE DIRECT ANALYSIS OF SOLID SAMPLES BY GRAPHITE-FURNACE ATOMIC-ABSORPTION SPECTROMETRY [J].
BAXTER, DC ;
FRECH, W .
FRESENIUS JOURNAL OF ANALYTICAL CHEMISTRY, 1990, 337 (03) :253-263
[4]   DETERMINATION OF METALS IN POLYVINYL-CHLORIDE) BY ATOMIC-ABSORPTION SPECTROMETRY .2. DETERMINATION OF LEAD AND MAGNESIUM IN SAMPLES OF POLYVINYL-CHLORIDE) WITH A HIGH CONTENT OF ALKALINE-EARTHS [J].
BELARRA, MA ;
ANZANO, JM ;
GALLARTA, F ;
CASTILLO, JR .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1987, 2 (01) :77-79
[5]   RAPID-DETERMINATION OF LEAD BY ANALYSIS OF SOLID SAMPLES USING GRAPHITE-FURNACE ATOMIC-ABSORPTION SPECTROMETRY [J].
BELARRA, MA ;
LAVILLA, I ;
ANZANO, JM ;
CASTILLO, JR .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1992, 7 (07) :1075-1078
[6]   SOLID SAMPLING IN ELECTROTHERMAL ATOMIC-ABSORPTION SPECTROMETRY USING COMMERCIAL ATOMIZERS - A REVIEW [J].
BENDICHO, C ;
DELOOSVOLLEBREGT, MTC .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1991, 6 (05) :353-374
[8]   ANALYSES OF SOLID SAMPLES BY GRAPHITE-FURNACE ATOMIC-ABSORPTION SPECTROMETRY USING ZEEMAN BACKGROUND CORRECTION [J].
CARNRICK, GR ;
LUMAS, BK ;
BARNETT, WB .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1986, 1 (06) :443-447
[9]   DIRECT DETERMINATION OF LEAD IN BOVINE LIVER BY SOLID SAMPLING WITH GRAPHITE-FURNACE ATOMIC-ABSORPTION SPECTROMETRY - A COMPARISON OF TUBE WALL ATOMIZATION, PLATFORM ATOMIZATION AND PROBE ATOMIZATION [J].
CHAKRABARTI, CL ;
KARWOWSKA, R ;
HOLLEBONE, BR ;
JOHNSON, PM .
SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 1987, 42 (11-12) :1217-1225
[10]   DIRECT ANALYSIS OF INORGANIC SOLIDS - EXPERIENCES OF A LABORATORY IN CEMENT INDUSTRIES [J].
ESSER, P .