Zeolite slurry analysis using freon-assisted graphite furnace vaporization for inductively coupled plasma atomic emission spectrometry

被引:17
作者
Alary, JF
Hernandez, G
Salin, ED
机构
[1] MCGILL UNIV,DEPT CHEM,MONTREAL,PQ H3A 2K6,CANADA
[2] UNIV HAVANA,INST MAT & REAGENTS ELECTR,HAVANA,CUBA
关键词
inductively coupled plasma atomic emission spectrometry; liquid analysis; zeolite; carbide-forming elements; halogen-assisted electrothermal vaporization; sheathing gas; slurry sample introduction; slurry injection;
D O I
10.1366/0003702953965948
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
Halogen-assisted electrothermal vaporization (ETV) was used as a sample introduction mechanism for inductively coupled plasma atomic emission spectrometry. Freon-12(TM) was used as a gas-phase halogenation reagent in the ETV graphite furnace to enhance vaporization of carbide-forming elements by converting them into metal halides. With Freon ETV, detection levels for Zn, Pb, Cd, Mn, V, Cu, Ti, and Ca were lower than those obtained with conventional nebulization of liquid standards. A zeolite sample was analyzed for the refractory elements chromium and vanadium. Without Freon, the signals for these elements were very low; however, with Freon, the signals were dramatically improved. Detection limits of 1.4 mu g/g and 0.3 mu g/g, respectively, were obtained for these elements. A sheathing and cooling gas flow was introduced to reduce analyte vapor condensation on tubing walls and to promote aggregate formation. This additional flow enhanced the quality of signal with the zeolite but proved to be detrimental for the determination of refractory elements in liquid standards with or without halogenation. The precision of determination for the zeolite sample varied from 3% for cadmium, to 6% for lead.
引用
收藏
页码:1796 / 1803
页数:8
相关论文
共 27 条
[1]  
BRECK DW, 1976, KIRKOTHMER ENCY CHEM, V15, P638
[2]  
BROEKAERT JAC, 1987, INDUCTIVELY COUPLE 1, P297
[3]  
DARKE SA, 1989, ANAL P, V26, P379
[4]   DETERMINATION OF IMPURITIES IN SILICON-CARBIDE POWDERS [J].
DOCEKAL, B ;
BROEKAERT, JAC ;
GRAULE, T ;
TSCHOPEL, P ;
TOLG, G .
FRESENIUS JOURNAL OF ANALYTICAL CHEMISTRY, 1992, 342 (1-2) :113-117
[5]  
HUE B, 1991, J ANAL ATOM SPECTROM, V6, P623
[6]  
HUE B, 1991, ANAL SCI, V7, P433
[7]   TRACE-ELEMENT ANALYSIS BY INDUCTIVELY COUPLED PLASMA ATOMIC EMISSION-SPECTROMETRY WITH ELECTROTHERMAL VAPORIZATION METHOD [J].
IDA, I ;
YOSHIKAWA, H ;
ISHIBASHI, Y ;
GUNJI, N .
ANALYTICAL SCIENCES, 1989, 5 (05) :615-618
[8]   INTERPRETING SOME ANALYTICAL CHARACTERISTICS OF THERMAL DISPERSION METHODS USED FOR SAMPLE INTRODUCTION IN ATOMIC SPECTROMETRY [J].
KANTOR, T .
SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 1988, 43 (9-11) :1299-1320
[10]   VOLATILIZATION OF SOME ELEMENTS FROM A GRAPHITE ROD DIRECT SAMPLE INSERTION DEVICE INTO AN INDUCTIVELY COUPLED ARGON PLASMA FOR OPTICAL-EMISSION SPECTROMETRY [J].
KIRKBRIGHT, GF ;
ZHANG, LX .
ANALYST, 1982, 107 (1275) :617-622