MONOCHROMATIC COMPUTED-TOMOGRAPHY OF THE HUMAN BRAIN USING SYNCHROTRON X-RAYS - TECHNICAL FEASIBILITY

被引:18
作者
NACHALIEL, E
DILMANIAN, FA
GARRETT, RF
THOMLINSON, WC
CHAPMAN, LD
GMUR, NF
LAZARZ, NM
MOULIN, HR
RIVERS, ML
RARBACK, H
STEFAN, PM
SPANNE, P
LUKE, PN
PEHL, R
THOMPSON, AC
MILLER, M
机构
[1] LAWRENCE BERKELEY LAB,BERKELEY,CA 94720
[2] ANAL CORP,PEABODY,MA 01960
关键词
D O I
10.1016/0168-9002(92)90570-T
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
A monochromatic computed tomography (CT) scanner is being developed at the X17 superconducting wiggler beamline at the National Synchrotron Light Source (NSLS), Brookhaven National Laboratory, to image the human head and neck. The system configuration is one of a horizontal fan beam and an upright seated rotating subject. The purposes of the project are to demonstrate improvement in the image contrast and in the image quantitative accuracy that can be obtained in monochromatic CT and to apply the system to specific clinical research programs in neuroradiology. This paper describes the first phantom studies carried out with a prototype system, using the dual photon absorptiometry (DPA) method at energies of 20 and 38 keV. The results show that improvements in image contrast and quantitative accuracy are possible with monochromatic DPA CT. Estimates of the clinical performance of the planned CT system are made on the basis of these initial results.
引用
收藏
页码:305 / 310
页数:6
相关论文
共 15 条
[1]   INTRAVENOUS ARTERIOGRAPHY USING SCANNED PROJECTION RADIOGRAPHY [J].
BRODY, WR ;
MACOVSKI, A ;
PELC, NJ ;
LEHMANN, L ;
JOSEPH, RA ;
EDELHEIT, LS .
RADIOLOGY, 1981, 141 (02) :509-514
[2]   PRECISE MEASUREMENT OF VERTEBRAL MINERAL-CONTENT USING COMPUTED-TOMOGRAPHY [J].
CANN, CE ;
GENANT, HK .
JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 1980, 4 (04) :493-500
[3]  
CHAPMAN LD, 1992, REV SCI INSTRUM, V63, P893
[4]   COMPUTED-TOMOGRAPHY WITH MONOCHROMATIC X-RAYS FROM THE NATIONAL-SYNCHROTRON-LIGHT-SOURCE [J].
DILMANIAN, FA ;
GARRETT, RF ;
THOMLINSON, WC ;
BERMAN, LE ;
CHAPMAN, LD ;
HASTINGS, JB ;
LUKE, PN ;
OVERSLUIZEN, T ;
SIDDONS, DP ;
SLATKIN, DN ;
STOJANOFF, V ;
THOMPSON, AC ;
VOLKOW, ND ;
ZEMAN, HD .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION B-BEAM INTERACTIONS WITH MATERIALS AND ATOMS, 1991, 56-7 :1208-1213
[5]  
DILMANIAN FA, 1990, PHYS MEDICA, V6, P301
[6]   A HIGH-ENERGY DOUBLE-CRYSTAL FIXED EXIT MONOCHROMATOR FOR THE X17 SUPERCONDUCTING WIGGLER BEAM LINE AT THE NSLS [J].
GARRETT, RF ;
DILMANIAN, FA ;
OVERSLUIZEN, T ;
LENHARD, A ;
BERMAN, LE ;
CHAPMAN, LD ;
STOEBER, W .
REVIEW OF SCIENTIFIC INSTRUMENTS, 1992, 63 (01) :595-598
[7]   X-RAY MONOCHROMATOR SYSTEM FOR USE WITH SYNCHROTRON RADIATION SOURCES [J].
GOLOVCHENKO, JA ;
LEVESQUE, RA ;
COWAN, PL .
REVIEW OF SCIENTIFIC INSTRUMENTS, 1981, 52 (04) :509-516
[8]  
Herman G.T., 1980, IMAGE RECONSTRUCTION
[9]   A MULTICHANNEL, HIGH LINEARITY CURRENT DIGITIZER FOR DIGITAL SUBTRACTION ANGIOGRAPHY [J].
NAKAMURA, M ;
KATZ, JE ;
THOMPSON, AC .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1988, 35 (01) :205-208
[10]   A FRAMEWORK FOR SPECTRAL ARTIFACT CORRECTIONS IN X-RAY CT [J].
STONESTROM, JP ;
ALVAREZ, RE ;
MACOVSKI, A .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1981, 28 (02) :128-141