A HIGH-RATE, LOW-NOISE, X-RAY SILICON STRIP DETECTOR SYSTEM

被引:27
作者
LUDEWIGT, B
JAKLEVIC, J
KIPNIS, I
ROSSINGTON, C
SPIELER, H
机构
[1] Lawrence Berkeley Laboratory, University of California, Berkeley, CA
关键词
D O I
10.1109/23.322854
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
An x-ray detector system, based on a silicon strip detector wire-bonded to a low noise charge-sensitive amplifier integrated circuit, has been developed for synchrotron radiation experiments which require very high count rates and good energy resolution. Noise measurements and x-ray spectra were taken using a 6 mm long, 55 mum pitch strip detector in conjunction with a prototype 16-channel charge-sensitive preamplifier, both fabricated using standard 1.2 mum CMOS technology. The detector system currently achieves an energy resolution of 350 eV FWHM at 5.9 keV, 2 mus peaking time, when cooled to -5-degrees-C.
引用
收藏
页码:1037 / 1041
页数:5
相关论文
共 12 条
[1]  
BUCHER JP, IN PRESS
[2]  
FISCHER DA, 1989, 6TH P NAT C SYNCHR R, P123
[3]   DESIGN AND PROTOTYPING OF THE FRONT END READOUT SYSTEM FOR THE CDF SILICON VERTEX DETECTOR [J].
HABER, C ;
CARITHERS, WC ;
ELY, RP ;
HOLLAND, S ;
KIRSTEN, F ;
KLEINFELDER, S ;
SPIELER, H ;
WESTER, WC ;
WONG, M ;
AMIDEI, D ;
HRYCYK, M ;
KEPHART, R ;
TKACEK, S ;
TOLLESTRUP, A ;
YAREMA, RJ ;
ZIMMERMAN, T ;
BEDESCHI, F ;
GALEOTTI, S ;
MENZIONE, A ;
PUNZI, G ;
RAFFAELLI, F ;
ZETTI, F .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1990, 37 (03) :1120-1126
[4]   PROPERTIES OF CMOS DEVICES AND CIRCUITS FABRICATED ON HIGH-RESISTIVITY, DETECTOR-GRADE SILICON [J].
HOLLAND, S .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1992, 39 (04) :809-813
[5]   A MONOLITHICALLY INTEGRATED DETECTOR-PREAMPLIFIER ON HIGH-RESISTIVITY SILICON [J].
HOLLAND, S ;
SPIELER, H .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1990, 37 (02) :463-468
[7]   A FLEXIBLE 128 CHANNEL SILICON STRIP DETECTOR INSTRUMENTATION INTEGRATED-CIRCUIT WITH SPARSE DATA READOUT [J].
KLEINFELDER, SA ;
CARITHERS, WC ;
ELY, RP ;
HABER, C ;
KIRSTEN, F ;
SPIELER, HG .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1988, 35 (01) :171-175
[8]  
MOSES WW, 1993, UNPUB IEEE T NUCL SC
[9]  
ROSSINGTON C, 1991, ADV X RAY ANAL, V34, P337
[10]   Large area, low capacitance Si(Li) detectors for high rate x-ray applications [J].
Rossington, C.S. ;
Fine, P.M. ;
Madden, N.W. .
IEEE Transactions on Nuclear Science, 1993, 40 (4 pt 1) :354-359