PRECISION AND ACCURACY STUDIES OF IMPACTION-ELECTROTHERMAL ATOMIZATION ATOMIC SPECTROMETRY

被引:6
作者
SNEDDON, J
机构
[1] Department of Chemistry, University of Lowell, Lowell
关键词
Accuracy impaction-Electrothermal Atomic Absorption Spectrometry; Precision;
D O I
10.1080/00032719008053449
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
The precision of aqueous solutions of cadmium, copper, and manganese introduced as an aerosol by an pneumatic nebulizer and electrothermal vaporizer to an impaction-electrothermal atoMization atomic absorption spectrometeric system was in the range of 1–3 % which was similar to that obtained by manual introduction of similar concentrations of aqueous solutions of cadmium, copper, and manganese to the electrothermal atomizer and atomic absorption spectrometry. The precision of a laboratory air sample was 7.6-9.9%. Accuracy was assessed by comparison to conventional methods of sampling air by collection on a filter followed by digestion of the filter and analysis by flame atomic absorption spectrometry The levels were found to be 59–69 % compared to the conventional method. © 1990, Taylor & Francis Group, LLC. All rights reserved.
引用
收藏
页码:1107 / 1112
页数:6