SPECTROFLUORIMETRIC DETERMINATION OF SULFATE IN WATERS IN NORMAL AND OPEN CLOSED FLOW-INJECTION CONFIGURATIONS

被引:12
作者
FERNANDEZBAND, B [1 ]
LINARES, P [1 ]
DECASTRO, MDL [1 ]
VALCARCEL, M [1 ]
机构
[1] UNIV CORDOBA,FAC SCI,DEPT ANALYT CHEM,E-14004 CORDOBA,SPAIN
关键词
SULFATE; SPECTROFLUORIMETRY; FLOW INJECTION; OPEN CLOSED SYSTEM; TERNARY COMPLEX;
D O I
10.1039/an9911600305
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A spectrofluorimetric method for the determination of sulphate, based on the formation of a ternary complex with biacetyl monoxime nicotinylhydrazone and zirconium is described. Two flow injection manifolds [normal (nFl) and open/closed] were optimized and used for the determination of sulphate in various water samples with good selectivity. The repeatability and linear range of the calibration graphs was better for the open/closed than for the nFl system (relative standard deviation 2.33% versus 3.51%, and linearity 1.5-150 mu-g ml-1 versus two linear portions of 2-30 and 30-150-mu-g ml-1), but the sampling rate using the nFl configuration (30 samples h-1) was at least twice that of the open/closed system.
引用
收藏
页码:305 / 307
页数:3
相关论文
共 23 条