SIMULTANEOUS MEASUREMENT OF ION RATIOS BY INDUCTIVELY-COUPLED PLASMA-MASS SPECTROMETRY WITH A TWIN-QUADRUPOLE INSTRUMENT

被引:19
作者
WARREN, AR
ALLEN, LA
PANG, HM
HOUK, RS
JANGHORBANI, M
机构
[1] IOWA STATE UNIV SCI & TECHNOL, US DOE, AMES LAB, DEPT CHEM, AMES, IA 50011 USA
[2] CTR STABLE ISOTOPE RES INC, CHICAGO, IL 60615 USA
关键词
INDUCTIVELY COUPLED PLASMA-MASS SPECTROMETRY; ICP-MS; ELEMENTAL ANALYSIS; ISOTOPE RATIOS;
D O I
10.1366/0003702944027958
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
The ion beam from the inductively coupled plasma (ICP) is split into two parts by a device comprised of two electrostatic analyzers in a back-to-back arrangement. Each ion beam is then monitored by a separate quadrupole mass analyzer and electron multiplier. Thus, ion ratios can be measured simultaneously without scanning. Isotope ratios for copper are measured with a precision of 0.14% relative standard deviation (RSD). Internal standard ratios for Co+/La+ are measured with a precision of 1% RSD. Ion signals in the two channels are highly correlated, and the magnitude of flicker noise is reduced in the resulting ratios. Bias in the measured isotope ratios is substantial but can be corrected by calibration.