ALTERED PHASE-ENCODING ORDER FOR REDUCED SENSITIVITY TO MOTION IN 3-DIMENSIONAL MR IMAGING

被引:34
作者
KORIN, HW
RIEDERER, SJ
BAMPTON, AEH
EHMAN, RL
机构
[1] Magnetic Resonance Laboratory, Mayo Clinic and Foundation, Rochester, Minnesota, 55905
来源
JMRI-JOURNAL OF MAGNETIC RESONANCE IMAGING | 1992年 / 2卷 / 06期
关键词
ARTIFACT; IMAGE PROCESSING; MOTION CORRECTION; PHASE-ENCODING ORDER; 3-DIMENSIONAL IMAGING;
D O I
10.1002/jmri.1880020613
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
A method of reordering phase and slab encoding that can be used to address some of the inherent problems due to motion in three dimensional imaging is described and implemented. The method is shown to be more robust with respect to reducing artifacts resulting from several fundamental types of motion. It can be readily implemented on a standard magnetic resonance imager with essentially no increase in total imaging time. Results of simulations and phantom and in vivo experiments are presented.
引用
收藏
页码:687 / 693
页数:7
相关论文
共 11 条
  • [1] Frahm J, Haase A, Matthaei D, Rapid three dimensional MR imaging using the FLASH technique, J Comput Assist Tomogr, 10, pp. 363-368, (1986)
  • [2] Johnson GA, Thompson MB, Drayer BP, Three‐dimensional MRI microscopy of the normal rat brain, Magn Reson Med, 4, pp. 351-365, (1987)
  • [3] Bailes DR, Gilderdale DJ, Bydder GM, Collins AG, Firmin DN, Respiratory ordered phase encoding (ROPE): a method for reducing respiratory motion artifacts in MR imaging, J Comput Assist Tomogr, 9, pp. 835-838, (1985)
  • [4] Korin HW, Felmlee JP, Ehman RL, Riederer SJ, Adaptive technique for three‐dimensional MR imaging of moving structures, Radiology, 177, pp. 217-221, (1990)
  • [5] Holsinger AE, Riederer SJ, The importance of phase‐encoding order in ultra‐short TR snapshot MR imaging, Magn Reson Med, 16, pp. 481-488, (1990)
  • [6] Mulkern RV, Wong STS, Winalski C, Jolesz FA, Contrast manipulation and artifact assessment of 2D and 3D RARE sequences, Magn Reson Imaging, 8, pp. 557-566, (1990)
  • [7] Holsinger-Bampton AE, Riederer SJ, Campeau NG, Ehman RL, Johnson CD, T1‐weighted snapshot gradient‐echo MR imaging of the abdomen, Radiology, 181, pp. 25-32, (1991)
  • [8] Korin HW, Ehman RL, Riederer SJ, Felmlee JP, Grimm RC, Respiratory kinematics of the upper abdominal organs: a quantitative study, Magn Reson Med, 23, pp. 172-178, (1992)
  • [9] Wood ML, Henkelman RM, MR image artifacts from periodic motion, Med Phys, 12, pp. 143-151, (1985)
  • [10] Mugler JP, Brookeman JR, Three‐dimensional magnetization‐prepared rapid gradient‐echo imaging (3D MP RAGE), Magn Reson Med, 15, pp. 152-157, (1990)