LONG-TERM FOLLOW-UP OF PATIENTS WITH SINGLE-LEAD VDD STIMULATION

被引:18
作者
FOLINO, AF
BUJA, G
RUZZA, L
NAVA, A
机构
[1] Department of Cardiology, University of Padova, Padova
来源
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY | 1994年 / 17卷 / 11期
关键词
SINGLE LEAD PACING; VDD PACING; FOLLOW-UP;
D O I
10.1111/j.1540-8159.1994.tb03762.x
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
The purpose of this study was to assess the long-term results of a single lead VDD pacing system (Phymos MPS) in 85 patients (48 males, 37 females, mean age 74 +/- 9 years). The system, which incorporates two atrial floating electrodes (Phymos 830-S), was implanted for second- or third-degree Mobitz II AV block. The percentage of AV Synchronized pacing, the presence of fusion beats due to short AV delay, and the occurrence of supraventricular arrhythmias was evaluated by Holter monitoring at 6-month intervals. Over a mean period of follow-up of 44 months (range 2-56), over 90% of AV synchronized pacing events were noted in 74 patients (87%). The presence of persistent fusion beats was detected in 12 patients (14%). In ten patients the pacemaker was reprogrammed to VVI mode for supraventricular arrhythmias (7 patients, 8.2%) or unsatisfactory atrial sensing (3 patients, 3.5%). One pacemaker was removed for pocket infection and two for battery depletion. These observations support the use of VDD single lead pacing to combine a physiological mode of stimulation with simple implantation techniques.
引用
收藏
页码:1854 / 1858
页数:5
相关论文
共 12 条
[1]   DDD PACEMAKERS MAXIMIZE HEMODYNAMIC BENEFITS AND MINIMIZE COMPLICATIONS FOR MOST PATIENTS [J].
BYRD, CL ;
SCHWARTZ, SJ ;
GONZALES, M ;
BYRD, CB ;
CIRALDO, RJ ;
SIVINA, M ;
YAHR, WZ ;
GREENBERG, JJ .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1988, 11 (11) :1911-1916
[2]  
CHAMBERLAIN DA, 1973, BRIT HEART J, V35, P559
[3]  
CURRY PVL, 1978, LANCET, V2, P757
[4]   A NEW ORTHOGONAL LEAD FOR P-SYNCHRONOUS PACING [J].
GOLDREYER, BN ;
OLIVE, AL ;
LESLIE, J ;
CANNOM, DS ;
WYMAN, MG .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1981, 4 (06) :638-644
[5]  
HASKELL RJ, 1989, BRIT HEART J, V61, P46
[6]   INCIDENCE OF ARRHYTHMIAS AFTER ATRIAL OR DUAL-CHAMBER PACEMAKER IMPLANTATION [J].
JORDAENS, L ;
ROBBENS, E ;
VANWASSENHOVE, E ;
CLEMENT, DL .
EUROPEAN HEART JOURNAL, 1989, 10 (02) :102-107
[7]  
KARLOF I, 1975, ACTA MED SCAND, V197, P195
[8]   A COMPARISON OF THE ACUTE AND LONG-TERM HEMODYNAMIC-EFFECTS OF VENTRICULAR INHIBITED AND ATRIAL SYNCHRONOUS VENTRICULAR INHIBITED PACING [J].
KRUSE, I ;
ARNMAN, K ;
CONRADSON, TB ;
RYDEN, L .
CIRCULATION, 1982, 65 (05) :846-855
[9]   OCCURRENCE AND SIGNIFICANCE OF ARRHYTHMIAS ASSOCIATED WITH ATRIAL-TRIGGERED VENTRICULAR PACING [J].
LEVANDERLINDGREN, M ;
PEHRSSON, SK .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1984, 7 (04) :628-639
[10]   EXPERIENCE AND IMPLANTATION TECHNIQUES WITH A NEW SINGLE-PASS LEAD VDD PACING SYSTEM [J].
LONGO, E ;
CATRINI, V .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1990, 13 (07) :927-936