We have developed a high-resolution ac susceptometer that uses a rf superconducting quantum interference device to directly measure the flux coupled into a superconducting detection coil from a sample's changing magnetic moment in an applied ac field. The system operates in a frequency range from 0.01 to 1500 Hz and an applied ac field range of 0.1-400 mu T with a sensitivity of about 5X10(-12) A m(2) for magnetic moment measurement, and at a reduced sensitivity down to 0.001 Hz. The instrument is based on an existing de magnetometer system and uses that system's temperature control and dc superconducting magnet to allow operation over a temperature range from 2 to 400 K and in applied dc fields of +/-5.0 T. During a measurement all operations are controlled automatically by computer from a menu-driven software system, with user input required only on initiation of a measurement sequence. Both real and imaginary components of the ac susceptibility can be determined.
引用
收藏
页码:2644 / 2652
页数:9
相关论文
共 15 条
[1]  
[Anonymous], MAGN PROP MEAS SYST
[2]   RADIO-FREQUENCY SUSCEPTIBILITY APPARATUS FOR MEASURING SMALL SUPERCONDUCTING SAMPLES [J].
DALRYMPLE, BJ ;
PROBER, DE .
REVIEW OF SCIENTIFIC INSTRUMENTS, 1984, 55 (06) :958-963
[3]   AUTOMATIC MUTUAL INDUCTANCE BRIDGE FOR ACCURATE AC SUSCEPTIBILITY MEASUREMENTS FROM 1.2 TO 300 K [J].
DEUTZ, AF ;
HULSTMAN, R ;
KRANENBURG, FJ .
REVIEW OF SCIENTIFIC INSTRUMENTS, 1989, 60 (01) :113-120
[4]   A SIMPLE DIGITAL SYSTEM FOR AC MAGNETIC MEASUREMENTS ON SUPERCONDUCTORS [J].
FABBRICATORE, P ;
GAMBARDELLA, U ;
GOMORY, F ;
MUSENICH, R ;
OCCHETTO, M ;
PARODI, R ;
POMPA, P .
REVIEW OF SCIENTIFIC INSTRUMENTS, 1991, 62 (07) :1796-1800
[5]   Principles and Methods of Low-Frequency Electric and Magnetic Measurements Using an rf-Biased Point-Contact Superconducting Device [J].
Giffard, R. P. ;
Webb, R. A. ;
Wheatley, J. C. .
JOURNAL OF LOW TEMPERATURE PHYSICS, 1972, 6 (5-6) :533-610
[6]  
Goldfarb, 1991, MAGNETIC SUSPECTIBIL, P49
[7]  
Hein R.A., 1991, MAGNETIC SUSCEPTIBIL
[8]  
HIANES DH, 1987, THESIS MONTANA STATE
[9]   VARIABLE FREQUENCY 2-PHASE AC SUSCEPTOMETER FOR USE IN DC BIAS FIELDS OVER THE TEMPERATURE-RANGE OF 4.2-K TO 300-K [J].
KOSZEGI, L ;
FOLDEAKI, M ;
DUNLAP, RA .
REVIEW OF SCIENTIFIC INSTRUMENTS, 1991, 62 (03) :793-796
[10]  
MULLER PH, 1982, J MAGN MAGN MATER, V28, P34