VENTILATOR TEMPERATURE SENSORS - AN UNUSUAL SOURCE OF PSEUDOMONAS-CEPACIA IN NOSOCOMIAL INFECTION

被引:27
作者
BERTHELOT, P
GRATTARD, F
MAHUL, P
JOSPE, R
POZZETTO, B
ROS, A
GAUDIN, OG
AUBOYER, C
机构
[1] UNIV ST ETIENNE, FAC MED J LISFRANC, DEPT MICROBIOL, F-42100 ST ETIENNE, FRANCE
[2] UNIV ST ETIENNE, FAC MED J LISFRANC, DEPT INTENS CARE, F-42100 ST ETIENNE, FRANCE
关键词
PSEUDOMONAS-CEPACIA; VENTILATORS; TEMPERATURE SENSORS; RIBOTYPING; NOSOCOMIAL INFECTION;
D O I
10.1016/0195-6701(93)90006-L
中图分类号
R1 [预防医学、卫生学];
学科分类号
1004 ; 120402 ;
摘要
A prospective study was undertaken to determine the source of Pseudomonas cepacia colonization and infection that had affected ventilated patients in an Intensive Care Unit (ICU) for three years. Thirty-eight patients undergoing mechanical ventilation were enrolled during a six-week period. Samples were taken from patients, ventilator circuits and the environment for culture. P. cepacia was isolated from the condensate formed in the ventilator circuit and the source of the contamination was shown to be the temperature sensor. Ribotyping of representative strains of P. cepacia performed with two endonucleases, EcoRI and PvuII, confirmed the homogeneity of the isolates from patients and ventilator circuits. A modification of the procedure for disinfection of the temperature sensors resulted in the eradication of P. cepacia from the ICU. © 1993.
引用
收藏
页码:33 / 43
页数:11
相关论文