HIGH-RESOLUTION REAL-TIME ULTRASONIC SCANNER

被引:52
作者
BERSON, M [1 ]
VAILLANT, L [1 ]
PATAT, F [1 ]
POURCELOT, L [1 ]
机构
[1] CHU TOURS,SERV DERMATOL,F-37033 TOURS,FRANCE
关键词
ULTRASONIC SCANNER; REAL TIME; HIGH RESOLUTION; HIGH FREQUENCY; DERMATOLOGY; NORMAL SKIN; TUMORS;
D O I
10.1016/0301-5629(92)90087-Q
中图分类号
O42 [声学];
学科分类号
070206 ; 082403 ;
摘要
High spatial resolution is required for echographic exploration of the skin, microvessels or small laboratory animals. With the scanner described here, high resolution is obtained by means of a strongly focused, wide-band 17 MHz center frequency transducer (-6 dB bandwidth: 22 MHz). The movement of this transducer above the skin provides a 6 mm wide and 5 mm deep echographic cross-section with an image rate of 15 images/s. The resolution is about 0.08 mm in axial and 0.2 to 0.3 mm in lateral directions. The device was tested on phantoms in water and in vivo on normal and pathological skin in the Department of Dermatology. With the easy-to-handle probe, explorations were made on psoriasis, basocellular carcinoma, malignant melanoma and sarcoidosis.
引用
收藏
页码:471 / 478
页数:8
相关论文
共 17 条
[1]   DETERMINING SKIN THICKNESS WITH PULSED ULTRA SOUND [J].
ALEXANDER, H ;
MILLER, DL .
JOURNAL OF INVESTIGATIVE DERMATOLOGY, 1979, 72 (01) :17-19
[2]  
BREITBART EW, 1985, BIENG SKIN, V1, P231
[3]  
DERIGAL J, 1989, J INVEST DERMATOL, V5, P621
[4]   HIGH-FREQUENCY ULTRASONIC-IMAGING OF SKIN - EXPERIMENTAL RESULTS [J].
DINES, KA ;
SHEETS, PW ;
BRINK, JA ;
HANKE, CW ;
CONDRA, KA ;
CLENDENON, JL ;
GOSS, SA ;
SMITH, DJ ;
FRANKLIN, TD .
ULTRASONIC IMAGING, 1984, 6 (04) :408-434
[5]  
ESCOFFIER C, 1986, BIOENG SKIN, V2, P87
[6]  
FREW HS, 1985, BIOENG SKIN, V1, P193
[7]   A 20-MHZ ULTRASOUND SYSTEM FOR IMAGING THE INTESTINAL WALL [J].
MARTIN, RW ;
SILVERSTEIN, FE ;
KIMMEY, MB .
ULTRASOUND IN MEDICINE AND BIOLOGY, 1989, 15 (03) :273-280
[8]   HUMAN-SKIN HISTOLOGY USING HIGH-RESOLUTION ECHOGRAPHY [J].
MURAKAMI, S ;
MIKI, Y .
JOURNAL OF CLINICAL ULTRASOUND, 1989, 17 (02) :77-82
[9]   INTRAVASCULAR US OF THE DISTAL AORTA AND ILIAC VESSELS - INITIAL FEASIBILITY STUDIES [J].
NISHIMURA, RA ;
WELCH, TJ ;
STANSON, AW ;
SHEEDY, PF ;
HOLMES, DR .
RADIOLOGY, 1990, 176 (02) :523-525
[10]   ULTRASOUND ANGIOSCOPY - REAL-TIME, TWO-DIMENSIONAL, INTRALUMINAL ULTRASOUND IMAGING OF BLOOD-VESSELS [J].
PANDIAN, NG ;
KREIS, A ;
BROCKWAY, B ;
ISNER, JM ;
SACHAROFF, A ;
BOLEZA, E ;
CARO, R ;
MULLER, D .
AMERICAN JOURNAL OF CARDIOLOGY, 1988, 62 (07) :493-494