DEVICE FOR MEASURING SOFT-TISSUE INTERFACE PRESSURES

被引:16
作者
BARBENEL, JC
SOCKALINGHAM, S
机构
[1] Bioengineering Unit, University of Strathclyde, Wolfson Centre, Glasgow
来源
JOURNAL OF BIOMEDICAL ENGINEERING | 1990年 / 12卷 / 06期
关键词
continuous output; interface pressure; Transducer;
D O I
10.1016/0141-5425(90)90062-R
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
This paper describes the construction and performance of a simple pressure sensing device with a continuous electrical output. It was constructed utilizing a commercially available transducer, an electropneumatic sensor capsule and a 1 m long tube. The transducer used was a piezo-resistive pressure-sensitive device producing an output voltage proportional to the applied pressure. This low cost, high accuracy device is temperature compensated and shows good linearity and negligible hysteresis. The sensor cell has a good thickness-to-diameter ratio and is sufficiently flexible to conform to most contours of the body. The tubing that conveys the pressure transmitting fluid also serves as a means of keeping the transducer distant from the measuring site. The device showed a highly satisfactory performance under laboratory conditions and has proven to be robust and reliable when used for clinical studies. © 1990.
引用
收藏
页码:519 / 522
页数:4
相关论文
共 10 条
[1]  
Bader D. L., 1985, Biomechanical measurement in orthopaedic practice, P145
[2]  
BARBENEL JC, 1983, PRESSURE SORES, P67
[3]  
Ferguson-Pell M. W., 1980, Engineering in Medicine, V9, P209, DOI 10.1243/EMED_JOUR_1980_009_052_02
[4]  
GUTTMANN L, 1976, BED SORE BIOMECHANIC, P153
[5]   ALTERATION OF HYPERTROPHIC SCARS INDUCED BY MECHANICAL PRESSURE [J].
KISCHER, CW ;
SHETLAR, MR ;
SHETLAR, CL .
ARCHIVES OF DERMATOLOGY, 1975, 111 (01) :60-64
[6]  
Mooney V, 1971, Bull Prosthet Res, V10, P129
[7]  
PATEL A, 1989, Journal of Rehabilitation Research and Development, V26, P55
[8]  
PIJKEREN TV, 1980, B PROSTHETICS RES DE, V17, P31
[9]  
1968, PRESSURE FORCE MEASU
[10]  
1989, DRUG TARIFF, P80