INVITRO INVESTIGATION OF THE FACTORS AFFECTING PULSE OXIMETRY

被引:18
作者
DEKOCK, JP
TARASSENKO, L
机构
[1] Medical Engineering Unit, Department of Engineering Science, Oxford University, Oxford
来源
JOURNAL OF BIOMEDICAL ENGINEERING | 1991年 / 13卷 / 01期
关键词
PULSE OXIMETRY; INVITRO MODEL; ARTERIAL OXYGEN SATURATION; MULTIPLE SCATTERING;
D O I
10.1016/0141-5425(91)90046-A
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The effect of a number of physiological parameters on pulse oximetry accuracy has been investigated in an in vitro model. We have found that above 50% saturation, pulse oximeters will not be affected by variations in haematocrit, blood flow rate, tissue blood content and pulse amplitude. At low saturations, however, it is known that the accuracy of pulse oximeters decreases and our in vitro results suggest how this may be corrected.
引用
收藏
页码:61 / 66
页数:6
相关论文
共 11 条
[1]  
KELLEHER JF, 1989, J CLIN MONITOR, V5, P50
[2]   AN INVITRO TISSUE MODEL FOR EVALUATING THE EFFECT OF CARBOXYHEMOGLOBIN CONCENTRATION ON PULSE OXIMETRY [J].
MENDELSON, Y ;
KENT, JC .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1989, 36 (06) :625-627
[3]   VARIATIONS IN OPTICAL-ABSORPTION SPECTRA OF ADULT AND FETAL HEMOGLOBINS AND ITS EFFECT ON PULSE OXIMETRY [J].
MENDELSON, Y ;
KENT, JC .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1989, 36 (08) :844-848
[4]  
Payne J. P., 1986, PULSE OXIMETRY
[5]   ACCURACY OF RESPONSE OF 6 PULSE OXIMETERS TO PROFOUND HYPOXIA [J].
SEVERINGHAUS, JW ;
NAIFEH, KH .
ANESTHESIOLOGY, 1987, 67 (04) :551-558
[6]   EFFECT OF ANEMIA ON PULSE OXIMETER ACCURACY AT LOW SATURATION [J].
SEVERINGHAUS, JW ;
KOH, SO .
JOURNAL OF CLINICAL MONITORING, 1990, 6 (02) :85-88
[7]   EFFECTS OF MULTIPLE-SCATTERING AND PERIPHERAL-CIRCULATION ON ARTERIAL OXYGEN-SATURATION MEASURED WITH A PULSE-TYPE OXIMETER [J].
SHIMADA, Y ;
YOSHIYA, I ;
OKA, N ;
HAMAGURI, K .
MEDICAL & BIOLOGICAL ENGINEERING & COMPUTING, 1984, 22 (05) :475-478
[8]   ROLE OF LIGHT-SCATTERING IN WHOLE-BLOOD OXIMETRY [J].
STEINKE, JM ;
SHEPHERD, AP .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1986, 33 (03) :294-301
[9]   PULSE OXIMETRY - ANALYSIS OF THEORY, TECHNOLOGY, AND PRACTICE [J].
WUKITSCH, MW ;
PETTERSON, MT ;
TOBLER, DR ;
POLOGE, JA .
JOURNAL OF CLINICAL MONITORING, 1988, 4 (04) :290-301
[10]   EVALUATION OF PULSE OXIMETRY [J].
YELDERMAN, M ;
NEW, W .
ANESTHESIOLOGY, 1983, 59 (04) :349-352