THE RIGHT VENTRICULAR OUTFLOW TRACT AS AN ALTERNATIVE PERMANENT PACING SITE - LONG-TERM FOLLOW-UP

被引:53
作者
BARIN, ES
JONES, SM
WARD, DE
CAMM, AJ
NATHAN, AW
机构
[1] ST BARTHOLOMEWS HOSP,DEPT CARDIOL,LONDON EC1A 7BE,ENGLAND
[2] BROMPTON HOSP,DEPT CARDIOL,LONDON SW3 6HP,ENGLAND
[3] ST GEORGE HOSP,LONDON,ENGLAND
[4] RYDE HOSP,SYDNEY,AUSTRALIA
来源
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY | 1991年 / 14卷 / 01期
关键词
VENTRICULAR OUTFLOW TRACT; PERMANENT PACING;
D O I
10.1111/j.1540-8159.1991.tb04040.x
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
The long-term characteristics of the right ventricular outflow tract have been assessed as an alternative permanent pacing site to the right ventricular apex. Thirty-three consecutive patients requiring ventricular pacing were randomized to be paced from one of the two sites. Pacing was performed using a screw-in lead, and a programmable pacemaker was used to facilitate threshold testing. There was no significant difference in the lead positioning time or any acute implant measurement (e.g., threshold at 0.5 msec 0.4 +/- 0.2 V for both sites, P = 0.99). Chronic measurements were also comparable during follow-up (mean 73 months) with a mean threshold at most recent follow-up of 0.15 +/- 0.2 msec (apex) and 0.13 +/- 0.21 msec (outflow tract) at 5 V, P = 0.81. There was only one pacing related complication, a lead dislodgement (outflow tract) in a pacemaker twiddler. Overall, both sites were highly satisfactory.
引用
收藏
页码:3 / 6
页数:4
相关论文
共 6 条
[1]  
BAYLISS CE, 1968, CAN MED ASSOC J, V99, P371
[2]   3-YEAR CLINICAL-EXPERIENCE WITH A NEW ENDOCARDIAL SCREW-IN LEAD WITH INTRODUCTION PROTECTION FOR USE IN THE ATRIUM AND VENTRICLE [J].
BISPING, HJ ;
KREUZER, J ;
BIRKENHEIER, H .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1980, 3 (04) :424-435
[3]   DETACHMENT OF PACEMAKER ACTIVE FIXATION HELIX [J].
DAWKINS, KD ;
JONES, S ;
WARD, DE .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1983, 6 (04) :818-819
[4]  
EIGEL P, 1981, PACE, V4, P85
[5]   TRANSVENOUS PACEMAKER INSERTION WITH A ZERO DISLODGEMENT RATE [J].
MORSE, D ;
YANKASKAS, M ;
JOHNSON, B ;
SPAGNA, P ;
LEMOLE, GM .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1983, 6 (02) :283-290
[6]  
SANTINI M, 1981, PACE, V4, P94