Environmental electromagnetic interference from electronic article surveillance devices: Interactions with an ICD

被引:24
作者
McIvor, ME
机构
[1] Heart Institute of St. Petersburg, St Petersburg, Florida
来源
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY | 1995年 / 18卷 / 12期
关键词
ICD; electronic article surveillance device;
D O I
10.1111/j.1540-8159.1995.tb04654.x
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
A patient with an implantable cardioverter defibrillator (ICD) experienced an inappropriate firing while in close proximity to an electronic article surveillance (EAS) device. Testing revealed the ICD was able to detect high frequency ''noise'' when close to the EAS device. ICD patients may need counseling to avoid close contact with such devices.