THE PERFORMANCE OF ULTRASOUND PHYSIOTHERAPY MACHINES IN LOTHIAN REGION, SCOTLAND, 1992

被引:36
作者
PYE, SD
MILFORD, C
机构
[1] Department of Medical Physics and Medical Engineering, Western General Hospital, Edinburgh, EH4 2XU, Crewe Road
关键词
ULTRASOUND; PHYSIOTHERAPY; CALIBRATION;
D O I
10.1016/0301-5629(94)90003-5
中图分类号
O42 [声学];
学科分类号
070206 ; 082403 ;
摘要
Eighty-five ultrasound therapy machines in use in Lothian Region, Scotland were tested for performance and their recalibration was expedited where necessary. The performance characteristics reported are the temporal average acoustic output power, the treatment frequency, pulse timing and treatment timer accuracy. Consideration is also given to the performance of the wide dynamic range radiation force balance that was used to make acoustic power measurements. Of the machines tested, 69% had power outputs that differed by more than 30% from the expected values. The results of power measurements are analysed according to continuous and pulsed output, single and dual frequency treatment heads and year of manufacture. Therapy machines more than 10-12 years old and modern dual frequency treatment heads performed particularly badly. This study suggests that the performance of ultrasound therapy machines has improved little in the last 20 years despite the introduction of IEC 601-2-5 in 1984.
引用
收藏
页码:347 / 359
页数:13
相关论文
共 24 条
[1]  
Allen K G, 1978, Physiotherapy, V64, P174
[2]   MINIMUM TARGET SIZE IN RADIATION FORCE MEASUREMENTS [J].
BEISSNER, K .
JOURNAL OF THE ACOUSTICAL SOCIETY OF AMERICA, 1984, 76 (05) :1505-1510
[3]   ON THE PLANE-WAVE APPROXIMATION OF ACOUSTIC INTENSITY [J].
BEISSNER, K .
JOURNAL OF THE ACOUSTICAL SOCIETY OF AMERICA, 1982, 71 (06) :1406-1411
[4]   THE SIGNIFICANCE OF MEMBRANE-CHANGES IN THE SAFE AND EFFECTIVE USE OF THERAPEUTIC AND DIAGNOSTIC ULTRASOUND [J].
DINNO, MA ;
DYSON, M ;
YOUNG, SR ;
MORTIMER, AJ ;
HART, J ;
CRUM, LA .
PHYSICS IN MEDICINE AND BIOLOGY, 1989, 34 (11) :1543-1552
[5]  
Dyson M, 1983, Ultrasound Med Biol, VSuppl 2, P61
[6]   THE BIOMECHANICAL EFFECTS OF LOW-INTENSITY ULTRASOUND ON HEALING TENDONS [J].
ENWEMEKA, CS ;
RODRIGUEZ, O ;
MENDOSA, S .
ULTRASOUND IN MEDICINE AND BIOLOGY, 1990, 16 (08) :801-807
[7]   THE IMPORTANCE OF MEASUREMENT OF EFFECTIVE TRANSDUCER RADIATING AREA IN THE TESTING AND CALIBRATION OF THERAPEUTIC ULTRASONIC INSTRUMENTS [J].
FYFE, MC ;
PARNELL, SM .
HEALTH PHYSICS, 1982, 43 (03) :377-381
[8]   DEVELOPMENT OF STANDARD MEASUREMENT METHODS FOR ESSENTIAL PROPERTIES OF ULTRASOUND THERAPY EQUIPMENT [J].
HEKKENBERG, RT ;
REIBOLD, R ;
ZEQIRI, B .
ULTRASOUND IN MEDICINE AND BIOLOGY, 1994, 20 (01) :83-98
[9]  
Hekkenberg RT, 1986, PHYSIOTHERAPY, V72, P390
[10]  
KOSSOFF G, 1962, ACUSTICA, V12, P84