DOSE REDUCTION DURING FLUOROSCOPIC PLACEMENT OF FEEDING TUBES

被引:17
作者
RUDIN, S
BEDNAREK, DR
MILLER, JA
机构
[1] Department of Radiology, Medicine/Biomed. Sciences Sch., State University of New York, Buffalo, NY 14215
关键词
DIAGNOSTIC RADIOLOGY; RADIATION EXPOSURE; FEEDING TUBE; RADIATIONS; EXPOSURE TO PATIENTS AND PERSONNEL;
D O I
10.1148/radiology.178.3.1994396
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
By both increasing the optical iris of the video camera and removing the grid in fluoroscopic procedures involved in placement of a Dobbhoff feeding tube, the radiation dose to patients and staff was reduced by five to seven times. An average expected dose to a patient of about 300 mGy per procedure was reduced by two to three times when the grid was removed and by an additional three times when an iris of increased diameter was used. Because Dobbhoff procedures do not involve obtaining a diagnosis, the image degradation was acceptable and did not affect the total exposure times or ability to conduct the procedure. With the grid out, the difference between the mean exposure times of 5.1 minutes in 96 patient studies done with a normal iris opening and 4.0 minutes in 52 studies done with an enlarged iris was not statistically significant. The importance of reducing patient dose is reinforced by the finding that one-third of the patients underwent repeated procedures, accounting for almost 60% of the total.
引用
收藏
页码:647 / 651
页数:5
相关论文
共 10 条
[1]   ASSESSMENT OF PATIENT EXPOSURE FOR BARIUM ENEMA EXAMINATIONS [J].
BEDNAREK, DR ;
RUDIN, S ;
WONG, R .
INVESTIGATIVE RADIOLOGY, 1983, 18 (05) :453-458
[2]   REDUCTION OF FLUOROSCOPIC EXPOSURE FOR THE AIR-CONTRAST BARIUM ENEMA [J].
BEDNAREK, DR ;
RUDIN, S ;
WONG, R ;
ANDRES, ML .
BRITISH JOURNAL OF RADIOLOGY, 1983, 56 (671) :823-828
[3]   THE GUT ORIGIN SEPTIC STATES IN BLUNT MULTIPLE TRAUMA (ISS = 40) IN THE ICU [J].
BORDER, JR ;
HASSETT, J ;
LADUCA, J ;
SEIBEL, R ;
STEINBERG, S ;
MILLS, B ;
LOSI, P ;
BORDER, D .
ANNALS OF SURGERY, 1987, 206 (04) :427-448
[4]   THE ELIMINATION OF GRIDS DURING INTENSIFIED FLUOROSCOPY AND PHOTOFLUORO SPOT IMAGING [J].
GRAY, JE ;
SWEE, RG .
RADIOLOGY, 1982, 144 (02) :426-429
[5]  
LEHMANN EL, 1975, NONPARAMETRICS
[6]   GASTROINTESTINAL FLUOROSCOPY - PATIENT DOSE AND METHODS FOR ITS REDUCTION [J].
LEIBOVIC, SJ ;
CALDICOTT, WJH .
BRITISH JOURNAL OF RADIOLOGY, 1983, 56 (670) :715-719
[7]  
Meyer S.L., 1975, DATA ANAL SCI ENG
[8]   VARIABLE APERTURE FLUOROSCOPIC UNIT FOR REDUCED PATIENT EXPOSURE [J].
ROSSI, RP ;
WESENBERG, RL ;
HENDEE, WR .
RADIOLOGY, 1978, 129 (03) :799-802
[9]   AN INVESTIGATION INTO THE LEVELS OF RADIATION EXPOSURE IN DIAGNOSTIC EXAMINATIONS INVOLVING FLUOROSCOPY [J].
ROWLEY, KA ;
HILL, SJ ;
WATKINS, RA ;
MOORES, BM .
BRITISH JOURNAL OF RADIOLOGY, 1987, 60 (710) :167-173
[10]   FLUOROSCOPIC PLACEMENT OF INTESTINAL TUBES [J].
SHIPPS, FC ;
SAYLER, CB ;
EGAN, JF ;
GREEN, GS ;
WEINSTEIN, CJ ;
JONES, JM .
RADIOLOGY, 1979, 132 (01) :226-227