A 2ND-ORDER SQUID GRADIOMETER OPERATING AT 77-K

被引:59
作者
TAVRIN, Y
ZHANG, Y
WOLF, W
BRAGINSKI, AI
机构
[1] Inst. fur Schicht- und Ionentechnik, Forschungszentrum Julich GmbH
关键词
D O I
10.1088/0953-2048/7/5/009
中图分类号
O59 [应用物理学];
学科分类号
摘要
We describe a portable second-order electronic SQUID gradiometer cooled with liquid nitrogen and capable of operating in unshielded space. The measured magnetic-field resolution is <300 fT Hz-1/2 and < 170 fT Hz-1/2 when operating in the first-order mode. The gradiometer was successfully used for the recording of screening magnetocardiograms of over 200 human subjects at four different geographic locations.
引用
收藏
页码:265 / 268
页数:4
相关论文
共 7 条
[1]   AN ELECTRONIC SECOND-ORDER GRADIOMETER FOR BIOMAGNETIC APPLICATIONS IN CLINICAL SHIELDED ROOMS [J].
Drung, D. ;
Koch, H. .
IEEE TRANSACTIONS ON APPLIED SUPERCONDUCTIVITY, 1993, 3 (01) :2594-2597
[2]  
ELBERT T, 1994, IN PRESS 9TH P INT C
[3]   DC SQUID MAGNETOMETERS FROM SINGLE LAYERS OF YBA2CU3O7-X [J].
KOELLE, D ;
MIKLICH, AH ;
LUDWIG, F ;
DANTSKER, E ;
NEMETH, DT ;
CLARKE, J .
APPLIED PHYSICS LETTERS, 1993, 63 (16) :2271-2273
[4]  
REIDEL P, THESIS TU CHEMNITZ G
[5]  
ROAS B, 1994, IN PRESS 1ST P EUR A
[6]   YBA2CU3O7 THIN-FILM SQUID GRADIOMETER FOR BIOMAGNETIC MEASUREMENTS [J].
TAVRIN, Y ;
ZHANG, Y ;
MUCK, M ;
BRAGINSKI, AI ;
HEIDEN, C .
APPLIED PHYSICS LETTERS, 1993, 62 (15) :1824-1826
[7]   A YBa2Cu3O7 THIN FILM SQUID GRADIOMETER FOR MEAS IN UNSHIELDED SPACE [J].
Tavrin, Y. ;
Zhang, Y. ;
Mueck, M. ;
Braginski, A. I. ;
Heiden, C. .
IEEE TRANSACTIONS ON APPLIED SUPERCONDUCTIVITY, 1993, 3 (01) :2477-2480