THE ESOPHAGEAL DETECTOR DEVICE - AN ASSESSMENT OF ACCURACY AND EASE OF USE BY PARAMEDICS

被引:12
作者
DONAHUE, PL [1 ]
机构
[1] UNIV MARYLAND,MED CTR,BALTIMORE,MD 21201
关键词
EQUIPMENT; ESOPHAGEAL DETECTOR DEVICE; INTUBATION; TRACHEAL; TRAINING;
D O I
10.1111/j.1365-2044.1994.tb04259.x
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
Accuracy, ease and speed of recognition of tracheal tube position were assessed using the oesophageal detector device in a series of 40 tests on 29 patients. A single blind method was used, with each paramedic performing a single test on each patient. The tests were randomly split between two groups consisting of those tests performed on the tracheal or oesophageal tube respectively. Of the 40 tests performed, 21 were on tracheal tubes and 19 on oesophageal tubes. Correct placement was diagnosed with 100% accuracy by the paramedics, none of whom had previously used the oesophageal detector device. Each paramedic also graded speed of recognition of position and ease of use of the device. Recognition of position was graded as instant in 37 out of 40 tests. Use of the oesophageal detector device by previously inexperienced paramedics has thus been shown to be accurate, rapid and easy to learn.
引用
收藏
页码:863 / 865
页数:3
相关论文
共 8 条
[1]  
BIRMINGHAM PK, 1986, ANESTH ANALG, V65, P886
[2]   EFFECTIVENESS OF USING END TIDAL CARBON-DIOXIDE CONCENTRATION TO MONITOR CARDIOPULMONARY RESUSCITATION [J].
HIGGINS, D ;
HAYES, M ;
DENMAN, W ;
WILKINSON, DJ .
BRITISH MEDICAL JOURNAL, 1990, 300 (6724) :581-581
[3]   END TIDAL CARBON-DIOXIDE DETECTOR FOR MONITORING CARDIOPULMONARY RESUSCITATION [J].
MUIR, JD ;
RANDALLS, PB ;
SMITH, GB .
BRITISH MEDICAL JOURNAL, 1990, 301 (6742) :41-42
[4]   THE ESOPHAGEAL DETECTOR DEVICE [J].
NUNN, JF .
ANAESTHESIA, 1988, 43 (09) :804-804
[5]  
O'CALLAGHAN J P, 1988, Canadian Journal of Anaesthesia, V35, pS59
[6]   ACCIDENTAL INTUBATION OF THE ESOPHAGUS [J].
POLLARD, BJ ;
JUNIUS, F .
ANAESTHESIA AND INTENSIVE CARE, 1980, 8 (02) :183-186
[7]   THE ESOPHAGEAL DETECTOR DEVICE - A PROSPECTIVE TRIAL ON 100 PATIENTS [J].
WILLIAMS, KN ;
NUNN, JF .
ANAESTHESIA, 1989, 44 (05) :412-414
[8]   THE ESOPHAGEAL DETECTOR DEVICE - DOES IT WORK [J].
ZALESKI, L ;
ABELLO, D ;
GOLD, MI .
ANESTHESIOLOGY, 1993, 79 (02) :244-247