DIAMOND-LIKE CARBON-FILMS FOR ENZYME ELECTRODES - CHARACTERIZATION OF NOVEL OVERLYING PERMSELECTIVE BARRIERS

被引:22
作者
HIGSON, SPJ [1 ]
VADGAMA, PM [1 ]
机构
[1] UNIV MANCHESTER,HOPE HOSP,DEPT MED,CLIN BIOCHEM SECT,SALFORD M6 8HD,LANCS,ENGLAND
关键词
BIOSENSORS; DIAMOND LIKE CARBON COATED FILMS; ENZYME ELECTRODES;
D O I
10.1016/0003-2670(94)00381-U
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Diamond like carbon (DLC) coated microporous polycarbonate membranes have been studied for use as novel composite permselective barriers membranes for a glucose enzyme electrode. Permeability coefficients, P, for key electrochemically active interferents across uncoated and DLC coated polycarbonate membranes has been compared. Interferent responses have then been assessed for sensors incorporating such membranes, and their relationship to differing DLC depositions assessed. Membranes with smaller pore sizes (0.03 and 0.01 mu m) and extended DLC depositions (up to 7 min coating), while imparting some enhanced selectivity towards glucose, failed to show major discrimination for glucose over interferents as shown by P values: maximum glucose-to-interferent P ratios being 1.36 and 1.25 for ascorbate and urate, respectively. The implications of these findings are discussed.
引用
收藏
页码:85 / 90
页数:6
相关论文
共 21 条
  • [1] Angus J. C., 1987, P EUROPEAN MATERIALS, V17, P179
  • [2] BARTLETT PN, 1991, J ELECTROANAL CHEM, V300, P75
  • [3] FERROCENE-MEDIATED ENZYME ELECTRODE FOR AMPEROMETRIC DETERMINATION OF GLUCOSE
    CASS, AEG
    DAVIS, G
    FRANCIS, GD
    HILL, HAO
    ASTON, WJ
    HIGGINS, IJ
    PLOTKIN, EV
    SCOTT, LDL
    TURNER, APF
    [J]. ANALYTICAL CHEMISTRY, 1984, 56 (04) : 667 - 671
  • [4] NEEDLE ENZYME ELECTRODES FOR BIOLOGICAL STUDIES
    CHURCHOUSE, SJ
    BATTERSBY, CM
    MULLEN, WH
    VADGAMA, PM
    [J]. BIOSENSORS, 1986, 2 (06) : 325 - 342
  • [5] EVANS AC, 1991, MED DEVICE TECHN MAY, P26
  • [6] FRANKS J, 1989, J VAC SCI A, V71, P2307
  • [7] A GLUCOSE ELECTRODE BASED ON CARBON PASTE CHEMICALLY MODIFIED WITH A FERROCENE-CONTAINING SILOXANE POLYMER AND GLUCOSE-OXIDASE, COATED WITH A POLY(ESTER-SULFONIC ACID) CATION-EXCHANGER
    GORTON, L
    KARAN, HI
    HALE, PD
    INAGAKI, T
    OKAMOTO, Y
    SKOTHEIM, TA
    [J]. ANALYTICA CHIMICA ACTA, 1990, 228 (01) : 23 - 30
  • [8] DIAMOND-LIKE CARBON COATED MICROPOROUS POLYCARBONATE AS A COMPOSITE BARRIER FOR A GLUCOSE ENZYME ELECTRODE
    HIGSON, SPJ
    VADGAMA, PM
    [J]. ANALYTICA CHIMICA ACTA, 1993, 271 (01) : 125 - 133
  • [9] ELECTRODE RESPONSES TO PHENOLIC SPECIES THROUGH CELLULOSIC MEMBRANES
    KOOCHAKI, Z
    CHRISTIE, I
    VADGAMA, P
    [J]. JOURNAL OF MEMBRANE SCIENCE, 1991, 57 (01) : 83 - 94
  • [10] GLUCOSE ENZYME ELECTRODE WITH EXTENDED LINEARITY - APPLICATION TO UNDILUTED BLOOD MEASUREMENTS
    MULLEN, WH
    KEEDY, FH
    CHURCHOUSE, SJ
    VADGAMA, PM
    [J]. ANALYTICA CHIMICA ACTA, 1986, 183 : 59 - 66