OPTIMIZATION OF HIGH-EFFICIENCY HEMODIALYSIS BY DETECTION AND CORRECTION OF FISTULA DYSFUNCTION

被引:92
作者
WINDUS, DW
AUDRAIN, J
VANDERSON, R
JENDRISAK, MD
PICUS, D
DELMEZ, JA
机构
[1] WASHINGTON UNIV,SCH MED,DEPT RADIOL & SURG,ST LOUIS,MO 63110
[2] WASHINGTON UNIV,CHROMALLOY AMER KIDNEY CTR,SCH MED,ST LOUIS,MO 63110
关键词
D O I
10.1038/ki.1990.206
中图分类号
R5 [内科学]; R69 [泌尿科学(泌尿生殖系疾病)];
学科分类号
1002 ; 100201 ;
摘要
Recirculation studies were performed in 103 patients treated with high-efficiency dialysis over a 14 month period. Fistulograms were performed on 22 out of 25 patients with greater than 0.15 fractional recirculation at a 400 ml/minute blood pump setting. Clinically significant abnormalities were found in 82% (N = 18) and treated in 17. Two patients had second episodes of elevated recirculations and were treated again within the period of follow-up. Treatment with angioplasty (N = 11) or surgical revision (N = 8) resulted in a fall in recirculation from 0.33 ± 0.04 to 0.12 ± 0.02 (P = 0.001). The fractional reduction of urea clearance due to recirculation fell from 0.20 ± 0.03 to 0.08 ± 0.02 (P = 0.001) and the effective urea clearance of the dialysis treatment rose by 16% from 193 ± 7 ml/min to 224 ± 6 ml/min (P = 0.001). Pre-dialysis BUN fell from 72 ± 4 mg/100 ml to 62 ± 3 mg/100 ml (P = 0.012). There was no correlation between venous pressure (VP) at 400 ml/min blood pump setting and recirculation (R2 = 0.04), although VP changed significantly comparing values before and after fistula repair (211 ± 10 vs. 186 ± 7 mmHg, P = 0.012). Venous pressures in 20 of the patients in our dialysis unit with recirculations of <0.10 were 201 ± 6 mmHg (P = NS compared to patients with recirculation ≥0.15 at 400 ml/min blood flow). We conclude that: a) significant reductions of fistula flow rates as determined by increased recirculation occurs in the absence of other abnormalities in fistula function such as increased VP or thrombosis; and b) correction of fistula abnormalities detected by increased recirculation improves the quality of the dialysis procedure by increasing effective clearance of solutes.
引用
收藏
页码:337 / 341
页数:5
相关论文
共 11 条
[1]   SEVERE DIALYZER DYSFUNCTION UNDETECTABLE BY STANDARD REPROCESSING VALIDATION TESTS [J].
DELMEZ, JA ;
WEERTS, CA ;
HASAMEAR, PD ;
WINDUS, DW .
KIDNEY INTERNATIONAL, 1989, 36 (03) :478-484
[2]   A MECHANISTIC ANALYSIS OF THE NATIONAL COOPERATIVE DIALYSIS STUDY (NCDS) [J].
GOTCH, FA ;
SARGENT, JA .
KIDNEY INTERNATIONAL, 1985, 28 (03) :526-534
[3]  
Ilstrup K, 1985, Trans Am Soc Artif Intern Organs, V31, P164
[4]  
JENDRISAK MD, IN PRESS ANN SURG
[5]  
Kobrin S M, 1989, ASAIO Trans, V35, P508, DOI 10.1097/00002480-198907000-00108
[6]   RECIRCULATION - REVIEW, TECHNIQUES FOR MEASUREMENT AND ABILITY TO PREDICT HEMOACCESS STENOSIS BEFORE AND AFTER ANGIOPLASTY [J].
NARDI, L ;
BOSCH, J .
BLOOD PURIFICATION, 1988, 6 (02) :85-89
[7]   USE OF DIGITAL SUBTRACTION ANGIOGRAPHY FOR EVALUATION OF VASCULAR ACCESS FOR HEMODIALYSIS [J].
PICUS, D ;
VANBREDA, A ;
KATZEN, BT ;
STEINBERG, DL .
CARDIOVASCULAR AND INTERVENTIONAL RADIOLOGY, 1987, 10 (04) :210-214
[8]  
SCHMIDT DF, 1990, KIDNEY INT, V37, P319
[9]   PREVENTION OF HEMODIALYSIS FISTULA THROMBOSIS - EARLY DETECTION OF VENOUS STENOSES [J].
SCHWAB, SJ ;
RAYMOND, JR ;
SAEED, M ;
NEWMAN, GE ;
DENNIS, PA ;
BOLLINGER, RR .
KIDNEY INTERNATIONAL, 1989, 36 (04) :707-711
[10]  
SHINABERGER JH, 1989, REPLACEMENT RENAL FU, pCH16