SCATTER FRACTIONS IN AMBER IMAGING

被引:20
作者
CHOTAS, HG [1 ]
FLOYD, CE [1 ]
DOBBINS, JT [1 ]
LO, JY [1 ]
RAVIN, CE [1 ]
机构
[1] DUKE UNIV,DEPT BIOMED ENGN,DURHAM,NC 27706
关键词
Lung; radiography; 60.1215; Mediastinum; 67.1215; Radiations; exposure to patients and personnel; Radiography; digital; Thorax;
D O I
10.1148/radiology.177.3.2244003
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Images of two phantoms were obtained with use of an advanced multiple-beam equalization radiography system, and scatter fractions were estimated with use of a photostimulable phosphor imaging system. Scatter fractions in the equalized images were lower in the mediastinum-equivalent areas and higher in the lung-equivalent areas, relative to images that were conventionally acquired with use of an antiscatter grid. The differences are attributed to a reduction in incident exposure in the lungs and the presence of cross-scatter between lung and mediastinal regions.