COMPUTED-TOMOGRAPHY WITH MONOCHROMATIC X-RAYS FROM THE NATIONAL-SYNCHROTRON-LIGHT-SOURCE

被引:30
作者
DILMANIAN, FA
GARRETT, RF
THOMLINSON, WC
BERMAN, LE
CHAPMAN, LD
HASTINGS, JB
LUKE, PN
OVERSLUIZEN, T
SIDDONS, DP
SLATKIN, DN
STOJANOFF, V
THOMPSON, AC
VOLKOW, ND
ZEMAN, HD
机构
[1] BROOKHAVEN NATL LAB,NATZ SYNCHROTRON LIGHT SOURCE,UPTON,NY 11973
[2] UNIV CALIF BERKELEY LAWRENCE BERKELEY LAB,BERKELEY,CA 94720
[3] UNIV SAO PAULO,INST PHYS,SAO PAULO,BRAZIL
关键词
D O I
10.1016/0168-583X(91)95133-X
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
A multiple-energy computed tomography (MECT) system that employs monochromatic and tunable 33-100 keV X rays from a superconducting wiggler at the National Synchrotron Light Source is being developed at Brookhaven National Laboratory. The CT configuration is that of a fixed, horizontal fan-shape beam and a subject seated in a rotating chair. Two quantitative CT methods will be used: a) K-edge subtraction of intravenously administered iodine (or a heavier element) to image brain tumors, large blood vessels of the lower head and neck, and arteriovenous malformations; and b) dual photon absorptiometry to obtain two brain CT images that map the low-Z elements and the intermediate-Z elements (i.e. P, S, Cl, K, Ca, and Fe) separately. The system is expected to provide 0.5 mm spatial resolution, horizontally, with unprecedented image contrast and accuracy of quantification. The system will employ a two-crystal monochromator and a high-purity Ge linear array detector.
引用
收藏
页码:1208 / 1213
页数:6
相关论文
共 33 条
[1]   INTRAVENOUS ARTERIOGRAPHY USING SCANNED PROJECTION RADIOGRAPHY [J].
BRODY, WR ;
MACOVSKI, A ;
PELC, NJ ;
LEHMANN, L ;
JOSEPH, RA ;
EDELHEIT, LS .
RADIOLOGY, 1981, 141 (02) :509-514
[2]   PRECISE MEASUREMENT OF VERTEBRAL MINERAL-CONTENT USING COMPUTED-TOMOGRAPHY [J].
CANN, CE ;
GENANT, HK .
JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 1980, 4 (04) :493-500
[3]   A BEAM-HARDENING CORRECTION USING DUAL-ENERGY COMPUTED-TOMOGRAPHY [J].
COLEMAN, AJ ;
SINCLAIR, M .
PHYSICS IN MEDICINE AND BIOLOGY, 1985, 30 (11) :1251-1256
[4]   A UHV COMPATIBLE 2-CRYSTAL MONOCHROMATOR FOR SYNCHROTRON RADIATION [J].
COWAN, PL ;
HASTINGS, JB ;
JACH, T ;
KIRKLAND, JP .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH, 1983, 208 (1-3) :349-353
[5]   A NEW SYSTEM FOR QUANTITATIVE ARTERIAL IMAGING AND BLOOD-FLOW MEASUREMENTS [J].
CUNNINGHAM, IA ;
HOBBS, BB ;
FENSTER, A .
INVESTIGATIVE RADIOLOGY, 1986, 21 (06) :465-471
[6]   QUANTITATIVE COMPUTED-TOMOGRAPHY IN ASSESSMENT OF OSTEOPOROSIS [J].
GENANT, HK ;
BLOCK, JE ;
STEIGER, P ;
GLUEER, CC ;
SMITH, R .
SEMINARS IN NUCLEAR MEDICINE, 1987, 17 (04) :316-333
[7]  
GMUR NF, 1989, BNL52205 REP
[8]   X-RAY MONOCHROMATOR SYSTEM FOR USE WITH SYNCHROTRON RADIATION SOURCES [J].
GOLOVCHENKO, JA ;
LEVESQUE, RA ;
COWAN, PL .
REVIEW OF SCIENTIFIC INSTRUMENTS, 1981, 52 (04) :509-516
[9]   OPTIMUM ENERGIES FOR X-RAY TRANSMISSION TOMOGRAPHY OF SMALL SAMPLES - APPLICATIONS OF SYNCHROTRON RADIATION TO COMPUTERIZED-TOMOGRAPHY .1. [J].
GRODZINS, L .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH, 1983, 206 (03) :541-545
[10]   PHOTOMASK TECHNIQUE FOR FABRICATING HIGH-PURITY GERMANIUM STRIP DETECTORS [J].
GUTKNECHT, D .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1990, 288 (01) :13-18