CLINICAL-EVALUATION OF THE TAKEDA MEDICAL (A-AND-D) TM-2420 AMBULATORY BLOOD-PRESSURE MONITOR - PRACTICAL EXPERIENCE AND COMPARISON WITH DIRECT AND INDIRECT MEASUREMENTS

被引:22
作者
HOEGHOLM, A
EIDEMAK, I
KRISTENSEN, KS
BANG, LE
MADSEN, NH
机构
[1] Department of Internal Medicine, County Central Hospital, Nastved
关键词
AMBULATORY BLOOD PRESSURE MONITORING; BLOOD PRESSURE MEASUREMENT; INTRAARTERIAL BLOOD PRESSURE;
D O I
10.3109/00365519209088357
中图分类号
R-3 [医学研究方法]; R3 [基础医学];
学科分类号
1001 ;
摘要
Takeda Medical (A & D) TM 2420 is an automatic ambulatory blood pressure monitoring system employing the auscultatory technique. The device was used under stable conditions and compared to readings from the Hawksley random-zero sphygmomanometer using a double headset stethoscope and a Y-connection. We tested 85 subjects (aged 13-89 years, systolic blood pressure 85-212 mmHg, diastolic blood pressure 40-116 mmHg) and found a difference amounting to 1.6+/-6.7 mmHg (mean+/-SD) for systolic and 2.1+/-4.5 mmHg for diastolic readings (Hawksley-TM 2420). In 62 subjects a comparison with simultaneous measurement on the opposite arm with the Hawksley manometer showed similar results. When comparing intra-arterial readings from 10 subjects, a difference (intra-arterial-TM 2420) of -1.9+/-12.1 mmHg was found for systolic pressures, while the diastolic difference was - 10.7+/-8.7 mmHg. Twenty-four hour monitoring was performed on 80 subjects; 70 of these yielded usable tracings. The proportion of successful recordings was acceptable, but the device was not suitable for bicycle stress testing. The quality of the accessories provided with the equipment could be improved, but in spite of this the monitoring system was found to be recommendable for clinical use.
引用
收藏
页码:261 / 268
页数:8
相关论文
共 14 条
[1]  
[Anonymous], 1985, J HYPERTENS, V3, P293
[2]  
BEVAN AT, 1969, CLIN SCI, V36, P329
[3]   STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1986, 1 (8476) :307-310
[4]   A VALIDATION-STUDY OF THE DEL MAR AVIONICS PRESSUROMETER-IV ACCORDING TO AAMI GUIDELINES [J].
HARSHFIELD, GA ;
HWANG, C ;
GRIM, CE .
JOURNAL OF HYPERTENSION, 1988, 6 (11) :913-918
[5]  
HOPE SL, 1988, J AMBULATORY MONITOR, V1, P39
[6]   BENCH AND AMBULATORY FIELD-EVALUATION OF THE A AND D TM-2420 AUTOMATED SPHYGMOMANOMETER [J].
JAMIESON, MJ ;
FOWLER, G ;
MACDONALD, TM ;
WEBSTER, J ;
WITTE, K ;
LAWSON, L ;
CRICHTON, W ;
JEFFERS, TA ;
PETRIE, JC .
JOURNAL OF HYPERTENSION, 1990, 8 (07) :599-605
[7]  
MAHESWARAN R, 1988, J AMBULAT MONIT, V1, P33
[8]   THE PROGNOSTIC VALUE OF AMBULATORY BLOOD PRESSURES [J].
PERLOFF, D ;
SOKOLOW, M ;
COWAN, R .
JAMA-JOURNAL OF THE AMERICAN MEDICAL ASSOCIATION, 1983, 249 (20) :2792-2798
[9]   OPTIMAL SIZE OF CUFF BLADDER FOR INDIRECT MEASUREMENT OF ARTERIAL-PRESSURE IN ADULTS [J].
RUSSELL, AE ;
WING, LMH ;
SMITH, SA ;
AYLWARD, PE ;
MCRITCHIE, RJ ;
HASSAM, RM ;
WEST, MJ ;
CHALMERS, JP .
JOURNAL OF HYPERTENSION, 1989, 7 (08) :607-613
[10]   EFFECT OF SIZE OF CUFF BLADDER ON ACCURACY OF MEASUREMENT OF INDIRECT BLOOD PRESSURE [J].
SIMPSON, JA ;
JAMIESON, G ;
DICKHAUS, DW ;
GROVER, RF .
AMERICAN HEART JOURNAL, 1965, 70 (02) :208-&