INSPIRATORY MUSCLE WORK OF BREATHING DURING FLOW-BY, DEMAND-FLOW, AND CONTINUOUS-FLOW SYSTEMS IN PATIENTS WITH CHRONIC OBSTRUCTIVE PULMONARY-DISEASE

被引:55
作者
SASSOON, CSH [1 ]
LODIA, R [1 ]
RHEEMAN, CH [1 ]
KUEI, JH [1 ]
LIGHT, RW [1 ]
MAHUTTE, CK [1 ]
机构
[1] UNIV CALIF IRVINE,IRVINE,CA 92717
来源
AMERICAN REVIEW OF RESPIRATORY DISEASE | 1992年 / 145卷 / 05期
关键词
D O I
10.1164/ajrccm/145.5.1219
中图分类号
R56 [呼吸系及胸部疾病];
学科分类号
摘要
The effect of flow-by continuous positive airway pressure (CPAP) in comparison with continuous-flow (CF) CPAP on inspiratory muscle work of breathing (WI) in intubated patients is not known. We hypothesized that WI during flow-by CPAP would be comparable with that during CF CPAR. In nine intubated male patients recovering from acute respiratory failure related to chronic obstructive pulmonary disease, we compared the effects of flow-by, demand-flow, and CF CPAP on WI. We also evaluated the extent to which the addition of 5 cm H2O of pressure support to demand-flow CPAP (DF-PS5) decreases WI. At CPAP levels of zero and 8 cm H2O, flow-by, demand-flow, DF-PS5 were applied in random order followed by CF CPAP for 15 min each. WI (expressed as Joules per liter and Joules per minute), maximal airway pressure drop during Inspiration (DELTA-Paw), total lung resistance (RL), lung compliance (CL), and ventilatory variables were measured. At CPAP of zero cm H2O, WI with flow-by was comparable with CF CPAP, and significantly less than with demand-flow CPAP, At both CPAP of zero and 8 cm H2O, the addition of 5 cm H2O PS to demand-flow CPAP reduced WI significantly to a level comparable with that of flow-by CPAP. At both CPAP levels, DELTA-Paw was the largest, with demand-flow CPAP. RL and CL were not significantly different between the different CPAP systems. We conclude that WI with flow-by CPAP is comparable with that with CF CPAP, and significantly less than with demand-flow CPAP. The addition of 5 cm H2O PS to demand-flow CPAP reduces WI to a level comparable with that of flow-by CPAP. The large DELTA-Paw, a result of the inability of the ventilator to meet the patient's early ventilatory demand, accounts for the greater WI during demand-flow CPAP.
引用
收藏
页码:1219 / 1222
页数:4
相关论文
共 17 条
[1]   INSPIRATORY WORK OF BREATHING DURING SPONTANEOUS VENTILATION USING DEMAND VALVES AND CONTINUOUS-FLOW SYSTEMS [J].
BEYDON, L ;
CHASSE, M ;
HARF, A ;
LEMAIRE, F .
AMERICAN REVIEW OF RESPIRATORY DISEASE, 1988, 138 (02) :300-304
[2]   INSPIRATORY PRESSURE SUPPORT PREVENTS DIAPHRAGMATIC FATIGUE DURING WEANING FROM MECHANICAL VENTILATION [J].
BROCHARD, L ;
HARF, A ;
LORINO, H ;
LEMAIRE, F .
AMERICAN REVIEW OF RESPIRATORY DISEASE, 1989, 139 (02) :513-521
[3]  
CAMPBELL EJM, 1970, RESPIRATORY MUSCLES, P115
[4]   COMPARISON OF WORK OF BREATHING ON HIGH GAS-FLOW AND DEMAND VALVE CONTINUOUS POSITIVE AIRWAY PRESSURE SYSTEMS [J].
GIBNEY, RTN ;
WILSON, RS ;
PONTOPPIDAN, H .
CHEST, 1982, 82 (06) :692-695
[5]   INTRINSIC PEEP AND ARTERIAL PCO2 IN STABLE PATIENTS WITH CHRONIC OBSTRUCTIVE PULMONARY-DISEASE [J].
HALUSZKA, J ;
CHARTRAND, DA ;
GRASSINO, AE ;
MILICEMILI, J .
AMERICAN REVIEW OF RESPIRATORY DISEASE, 1990, 141 (05) :1194-1197
[6]  
HIRSCH C A, 1988, Respiratory Care, V33, P908
[7]   INSPIRATORY WORK AND AIRWAY PRESSURE WITH CONTINUOUS POSITIVE AIRWAY PRESSURE DELIVERY SYSTEMS [J].
KATZ, JA ;
KRAEMER, RW ;
GJERDE, GE .
CHEST, 1985, 88 (04) :519-526
[8]  
PESLIN R, 1983, B EUR PHYSIOPATHO S5, V19, P33
[9]   CONTINUOUS POSITIVE AIRWAY PRESSURE REDUCES WORK OF BREATHING AND DYSPNEA DURING WEANING FROM MECHANICAL VENTILATION IN SEVERE CHRONIC OBSTRUCTIVE PULMONARY-DISEASE [J].
PETROF, BJ ;
LEGARE, M ;
GOLDBERG, P ;
MILICEMILI, J ;
GOTTFRIED, SB .
AMERICAN REVIEW OF RESPIRATORY DISEASE, 1990, 141 (02) :281-289
[10]   CALIBRATION OF RESPIRATORY INDUCTIVE PLETHYSMOGRAPH DURING NATURAL BREATHING [J].
SACKNER, MA ;
WATSON, H ;
BELSITO, AS ;
FEINERMAN, D ;
SUAREZ, M ;
GONZALEZ, G ;
BIZOUSKY, F ;
KRIEGER, B .
JOURNAL OF APPLIED PHYSIOLOGY, 1989, 66 (01) :410-420