IMPROVED FLOW CYTOMETRIC METHOD FOR HLA-B27 TYPING

被引:11
作者
JANSSEN, WCM [1 ]
ROUWEN, JACN [1 ]
HOFFMANN, JJML [1 ]
机构
[1] CATHARINA HOSP,DEPT CLIN LABS,POB 1350,5602 ZA EINDHOVEN,NETHERLANDS
关键词
IMMUNOFLUORESCENCE; LYMPHOCYTES; HUMAN LEUKOCYTE ANTIGEN;
D O I
10.1177/000456329202900608
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
HLA-B27 is a cell marker of clinical interest because of its high association with certain diseases. The HLA-B27 antigen was detected on lymphocytes using a monoclonal antibody in an indirect immuno-fluorescence assay using a fluorescence flow cytometer. The considerable crossreaction of the monoclonal antibody with the HLA-B7 antigen was effectively suppressed by masking it by means of human anti-HLA-B7 antiserum. The flow cytometric method was evaluated by comparing the results with those obtained by the standard lymphocytotoxicity test and showed complete agreement in 107 selected patient samples.
引用
收藏
页码:663 / 667
页数:5
相关论文
共 6 条
[1]  
ALBRECHT J, 1987, CLIN CHEM, V33, P1619
[2]  
BOYUM A, 1968, SCAND J CLIN LAB S97, V21, P1
[3]  
FIZET D, 1989, ANN BIOL CLIN-PARIS, V47, P408
[4]   HLA AND DISEASE 1982 - A SURVEY [J].
SVEJGAARD, A ;
PLATZ, P ;
RYDER, LP .
IMMUNOLOGICAL REVIEWS, 1983, 70 :193-218
[5]   MICRODROPLET ASSAY OF HUMAN SERUM CYTOTOXINS [J].
TERASAKI, PI ;
MCCLELLAND, JD .
NATURE, 1964, 204 (496) :998-&
[6]   DESCRIPTION OF A MOUSE MONOCLONAL ANTI-HLA-B27 ANTIBODY HLA-ABC-M3 [J].
TRAPANI, JA ;
VAUGHAN, HA ;
SPARROW, RL ;
TAIT, BD ;
MCKENZIE, IFC .
HUMAN IMMUNOLOGY, 1983, 7 (04) :205-216