SUBMICROMETER AEROSOL GENERATOR DEVELOPMENT FOR THE UNITED-STATES-ENVIRONMENTAL-PROTECTION-AGENCY HUMAN EXPOSURE LABORATORY

被引:6
作者
PETERS, TM
CHEIN, HM
LUNDGREN, DA
BERNTSEN, J
机构
[1] UNIV FLORIDA,DEPT ENVIRONM ENGN,GAINESVILLE,FL 32611
[2] TRC ENVIRONM CORP,CHAPEL HILL,NC 27514
关键词
D O I
10.1080/02786829408959663
中图分类号
TQ [化学工业];
学科分类号
0817 ;
摘要
A submicrometer aerosol generation unit was developed for use in the U.S. Environmental Protection Agency's new air pollution exposure chambers located on the campus of the University of North Carolina at Chapel Hill. The aerosol mass median aerodynamic diameter and mass concentration are operator-selected, ranging from 0.1 to 1.0 mum and from 0 to 2 mg/m3, respectively. The aerosol was to have a geometric standard deviation (GSD) of < 1.5. Two-fluid nebulization was selected to generate water-soluble aerosols. Multiple Collison-type nozzles are used to create a polydisperse droplet cloud. Aerosol uniformity is improved by removing the large droplets from the original distribution by passing the aerosol through an external impaction device. Variations in the solvent-solute ratio, nebulizer driving pressure, impactor cut-point size, and number of nozzles are used to control the final aerosol size and mass concentration. Aerosol size distributions were determined with an electrical aerosol analyzer and a low-pressure impactor. By means of data obtained by testing several experimental nebulizers, a 48-nozzle prototype submicrometer aerosol generator was designed, constructed, and tested. The GSD of the aerosol produced ranged from 1.4 to 1.6 and the mass output, upon generator replication, can satisfy maximum exposure chamber demands.
引用
收藏
页码:51 / 61
页数:11
相关论文
共 13 条
[1]   GENERATION OF SUBMICRON MASS VOLUME-MONODISPERSE AEROSOLS WITH A NEBULIZER-IMPACTOR-ELECTROSTATIC CLASSIFIER SYSTEM [J].
ASGHARIAN, B ;
FANG, CP ;
COHEN, BS .
AEROSOL SCIENCE AND TECHNOLOGY, 1992, 16 (02) :96-104
[2]   ATOMIZER OF HIGH-STABILITY [J].
COOPER, G .
AMERICAN INDUSTRIAL HYGIENE ASSOCIATION JOURNAL, 1979, 40 (08) :734-737
[3]   USE OF THE ELECTROSTATIC CLASSIFICATION METHOD TO SIZE 0.1 MU-M SRM PARTICLES - A FEASIBILITY STUDY [J].
KINNEY, PD ;
PUI, DYH ;
MULHOLLAND, GW ;
BRYNER, NP .
JOURNAL OF RESEARCH OF THE NATIONAL INSTITUTE OF STANDARDS AND TECHNOLOGY, 1991, 96 (02) :147-176
[4]   AN IMPROVED CONSTANT OUTPUT ATOMIZER [J].
LEONG, KH ;
WANG, HC ;
STUKEL, JJ ;
HOPKE, PK .
AMERICAN INDUSTRIAL HYGIENE ASSOCIATION JOURNAL, 1982, 43 (02) :135-136
[5]   AEROSOL GENERATOR OF HIGH STABILITY [J].
LIU, BYH ;
LEE, KW .
AMERICAN INDUSTRIAL HYGIENE ASSOCIATION JOURNAL, 1975, 36 (12) :861-865
[6]   IMPACTOR DESIGN [J].
MARPLE, VA ;
WILLEKE, K .
ATMOSPHERIC ENVIRONMENT, 1976, 10 (10) :891-896
[7]  
MAY KR, 1973, AEROSOL SCI, V4, P235
[8]   OPERATING CHARACTERISTICS OF SOME COMPRESSED-AIR NEBULIZERS [J].
MERCER, TT ;
TILLERY, MI ;
CHOW, HY .
AMERICAN INDUSTRIAL HYGIENE ASSOCIATION JOURNAL, 1968, 29 (01) :66-+
[9]   COMPARISON AND COMBINATION OF AEROSOL-SIZE DISTRIBUTIONS MEASURED WITH A LOW-PRESSURE IMPACTOR, DIFFERENTIAL MOBILITY PARTICLE SIZER, ELECTRICAL AEROSOL ANALYZER, AND AERODYNAMIC PARTICLE SIZER [J].
PETERS, TM ;
CHEIN, HM ;
LUNDGREN, DA ;
KEADY, PB .
AEROSOL SCIENCE AND TECHNOLOGY, 1993, 19 (03) :396-405
[10]   CHARACTERISTICS OF A DOUBLE-ORIFICE NEBULIZER [J].
PILACINSKI, W ;
SZEWCZYK, KW ;
LEHTIMAKI, M ;
WILLEKE, K .
JOURNAL OF AEROSOL SCIENCE, 1990, 21 (07) :977-982