PRECISION FIELD AVERAGING NMR MAGNETOMETER FOR LOW AND HIGH FIELDS, USING FLOWING WATER

被引:20
作者
PENDLEBURY, JM
SMITH, K
UNSWORTH, P
GREENE, GL
MAMPE, W
机构
[1] HARVARD UNIV,CAMBRIDGE,MA 02138
[2] INST MAX VON LAUE PAUL LANGEVIN,F-38042 GRENOBLE,FRANCE
关键词
D O I
10.1063/1.1135904
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
The separated oscillatory field magnetic resonance technique (Ramsey technique) has been employed with flowing water as a volume averaging magnetometer. Polarization and detection were performed in high fields, external to the volume over which the magnetic field was to be averaged. An accuracy of a few parts in 107 at a nominal field of 18 G was obtained. The technique is applicable, using standard equipment, for the measurement of fields ranging from kilogauss down to milligauss.
引用
收藏
页码:535 / 540
页数:6
相关论文
共 19 条
[1]  
BENEDEK G, 1952, COMMUNICATION
[2]   MOLECULAR-BEAM MAGNETIC RESONANCE STUDIES OF HD AND D2 [J].
CODE, RF ;
RAMSEY, NF .
PHYSICAL REVIEW A, 1971, 4 (05) :1945-&
[3]   MAGNETIC MOMENT OF THE NEUTRON [J].
COHEN, VW ;
CORNGOLD, NR ;
RAMSEY, NF .
PHYSICAL REVIEW, 1956, 104 (02) :283-291
[4]  
CORNGOLD N, 1955, THESIS HARVARD U
[5]   NEW MEASUREMENT OF MAGNETIC-MOMENT OF NEUTRON [J].
GREENE, GL ;
RAMSEY, NF ;
MAMPE, W ;
PENDLEBURY, JM ;
SMITH, K ;
DRESS, WD ;
MILLER, PD ;
PERRIN, P .
PHYSICS LETTERS B, 1977, 71 (02) :297-300
[6]  
GREENE GL, 1977, THESIS HARVARD U
[7]   DEMAGNETIZING FACTORS OF THE GENERAL ELLIPSOID [J].
OSBORN, JA .
PHYSICAL REVIEW, 1945, 67 (11-1) :351-357
[8]   MAGNETIC-MOMENT OF PROTON IN H2O IN BOHR MAGNETONS [J].
PHILLIPS, WD ;
COOKE, WE ;
KLEPPNER, D .
PHYSICAL REVIEW LETTERS, 1975, 35 (24) :1619-1622
[9]  
POPLE, 1959, HIGH RESOLUTION NMR
[10]  
PURCELL EM, 1952, COMMUNICATION