EFFECTS OF X-RAY-BEAM EQUALIZATION ON MAMMOGRAPHIC IMAGING

被引:27
作者
LAM, KL [1 ]
CHAN, HP [1 ]
机构
[1] UNIV CHICAGO,DEPT RADIOL,KURT ROSSMANN LABS RADIOL IMAGE RES,CHICAGO,IL 60637
关键词
D O I
10.1118/1.596502
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
We employed an elastic reservoir filled with tissue-equivalent fluid as an x-ray beam filter for exposure equalization. The filter is designed to be placed between the upper compression plate and the breast. During compression, it will conform to the shape of the peripheral region of the breast and fill the space between the upper and lower compression plates, thereby reducing the variation in x-ray penetration between the thick and thin regions of the breast and compressing the dynamic range of the breast image. Results of a phantom study showed that the visibility of low-contrast features was improved in the peripheral region of the equalized breast phantom image. Experimental measurements indicated that the filter reduced scattered radiation in the peripheral region. This result was confirmed by a Monte Carlo simulation study. The radiation contrast, with the effect of beam hardening included, was also improved when equalization was employed. © 1990, American Association of Physicists in Medicine. All rights reserved.
引用
收藏
页码:242 / 249
页数:8
相关论文
共 21 条
[1]   CONTRAST THRESHOLDS OF THE HUMAN EYE [J].
BLACKWELL, HR .
JOURNAL OF THE OPTICAL SOCIETY OF AMERICA, 1946, 36 (11) :624-643
[2]   THE VALIDITY OF MONTE-CARLO SIMULATION IN STUDIES OF SCATTERED RADIATION IN DIAGNOSTIC-RADIOLOGY [J].
CHAN, HP ;
DOI, K .
PHYSICS IN MEDICINE AND BIOLOGY, 1983, 28 (02) :109-129
[3]  
CHAN HP, 1981, THESIS U CHICAGO CHI
[4]  
CHAN HP, 1908, PHYS MED BIOL, V28, P565
[5]  
CURRY TS, 1984, CHRISTENSENS INTRO P, P322
[6]  
Fewell T. R., 1978, HDB MAMMOGRAPHIC XRA, P59
[7]  
HASEGAWA BH, 1986, RADIOLOGY, V159, P237
[8]   X-RAY SENSITOMETER FOR SCREEN-FILM COMBINATIONS USED IN MEDICAL RADIOLOGY [J].
HAUS, AG ;
ROSSMANN, K .
RADIOLOGY, 1970, 94 (03) :673-&
[9]  
KRUGER RA, 1984, BASIC CONCEPTS DIGIT, P183
[10]   OPTICAL-IMAGE PROCESSING WITH LIQUID-CRYSTAL DISPLAY FOR IMAGE INTENSIFIER TELEVISION SYSTEMS [J].
LAM, KL ;
CHAN, HP ;
DOI, K .
MEDICAL PHYSICS, 1988, 15 (06) :838-845