IN-VIVO DEGRADATION AND BIOCOMPATIBILITY STUDY OF IN-VITRO PRE-DEGRADED AS-POLYMERIZED POLYLACTIDE PARTICLES

被引:169
作者
BERGSMA, JE
ROZEMA, FR
BOS, RRM
BOERING, G
DEBRUIJN, WC
PENNINGS, AJ
机构
[1] ERASMUS UNIV ROTTERDAM,INST CLIN PATHOL 1,AEM UNIT,3000 DR ROTTERDAM,NETHERLANDS
[2] UNIV GRONINGEN,DEPT POLYMER CHEM,9700 RB GRONINGEN,NETHERLANDS
关键词
BIODEGRADATION; BIOCOMPATIBILITY; IN VITRO PRE-DEGRADATION; AS-POLYMERIZED POLYLACTIDES; ANIMAL MODEL;
D O I
10.1016/0142-9612(95)93253-A
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The degradation of high molecular weight as-polymerized poly(L-lactide) (PLLA) is very slow; it takes more than 5.6 yr for total resorption. Moreover, the degradation products of as-polymerized PLLA bone plates, consisting of numerous stable particles of high crystallinity, are related with a subcutaneous swelling in patients 3 yr postoperatively. In order to avoid these complications, polymers were developed that are anticipated to have comparable mechanical properties but a higher degradation rate and do not degrade into highly stable particles that can induce a subcutaneous swelling. On chemical grounds it can be expected that copolymerization of PLLA with 4% D-lactide (PLA96) or by modifying PLLA through cross-linking (CL-PLLA) will lead to less stable particles and a higher degradation rate. To evaluate the long-term suitability of these as-polymerized polymers, the biocompatibility of the degradation products should be studied. Considering the very slow degradation rate of as-polymerized PLLA, in vitro pre-degradation at elevated temperatures was used to shorten the in vivo follow-up periods. In this study, the biocompatibility and degradation of as-polymerized PLLA, PLA96 and CL-PLLA were investigated by implanting pre-degraded particulate materials subcutaneously in rats. Animals were killed after a postoperative period varying from 3 to 80 wk. Light and electron microscopical analysis and quantitative measurements were performed. The histological response of all three pre-degraded materials showed a good similarity with in vivo implanted material. Pre-degraded PLLA induced a mild foreign body reaction and showed a slow degradation rate. PLA96 and CL-PLLA had a substantially lower crystallinity, a smaller mean particle size and an enhanced degradation rate compared to PLLA. Based on the chemical and quantitative analysis, the degradation of PLA96 and CL-PLLA was much more enhanced and thus more favourable than the degradation of PLLA.
引用
收藏
页码:267 / 274
页数:8
相关论文
共 34 条
[1]   QUANTITATIVE CHARACTERIZATION OF CELLS AT THE INTERFACE OF LONG-TERM IMPLANTS OF SELECTED POLYMERS [J].
BEHLING, CA ;
SPECTOR, M .
JOURNAL OF BIOMEDICAL MATERIALS RESEARCH, 1986, 20 (05) :653-666
[2]   FOREIGN-BODY REACTIONS TO RESORBABLE POLY(L-LACTIDE) BONE PLATES AND SCREWS USED FOR THE FIXATION OF UNSTABLE ZYGOMATIC FRACTURES [J].
BERGSMA, EJ ;
ROZEMA, FR ;
BOS, RRM ;
DEBRUIJN, WC .
JOURNAL OF ORAL AND MAXILLOFACIAL SURGERY, 1993, 51 (06) :666-670
[3]  
BERGSMA JE, 1994, IN PRESS BIOMATERIAL
[4]   DEGRADATION OF AND TISSUE REACTION TO BIODEGRADABLE POLY(L-LACTIDE) FOR USE AS INTERNAL-FIXATION OF FRACTURES - A STUDY IN RATS [J].
BOS, RRM ;
ROZEMA, FR ;
BOERING, G ;
NIJENHUIS, AJ ;
PENNINGS, AJ ;
VERWEY, AB ;
NIEUWENHUIS, P ;
JANSEN, HWB .
BIOMATERIALS, 1991, 12 (01) :32-36
[5]   RESORBABLE POLY(L-LACTIDE) PLATES AND SCREWS FOR THE FIXATION OF ZYGOMATIC FRACTURES [J].
BOS, RRM ;
BOERING, G ;
ROZEMA, FR ;
LEENSLAG, JW .
JOURNAL OF ORAL AND MAXILLOFACIAL SURGERY, 1987, 45 (09) :751-753
[6]   BONE-PLATES AND SCREWS OF BIOABSORBABLE POLY (L-LACTIDE) - AN ANIMAL PILOT-STUDY [J].
BOS, RRM ;
ROZEMA, FR ;
BOERING, G ;
NIJENHUIS, AJ ;
PENNINGS, AJ ;
JANSEN, HWB .
BRITISH JOURNAL OF ORAL & MAXILLOFACIAL SURGERY, 1989, 27 (06) :467-476
[7]   DEGRADATION AND TISSUE REPLACEMENT OF AN ABSORBABLE POLYGLYCOLIDE SCREW IN THE FIXATION OF RABBIT FEMORAL OSTEOTOMIES [J].
BOSTMAN, O ;
PAIVARINTA, U ;
PARTIO, E ;
VASENIUS, J ;
MANNINEN, M ;
ROKKANEN, P .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1992, 74A (07) :1021-1031
[8]   POLYMERIC DEBRIS FROM ABSORBABLE POLYGLYCOLIDE SCREWS AND PINS - INTRAOSSEOUS MIGRATION STUDIED IN RABBITS [J].
BOSTMAN, O ;
PAIVARINTA, U ;
MANNINEN, M ;
ROKKANEN, P .
ACTA ORTHOPAEDICA SCANDINAVICA, 1992, 63 (05) :555-559
[9]   OSTEOLYTIC CHANGES ACCOMPANYING DEGRADATION OF ABSORBABLE FRACTURE FIXATION IMPLANTS [J].
BOSTMAN, OM .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1991, 73 (04) :679-682
[10]   THE BIODEGRADABILITY OF POLYESTER BLENDS [J].
CHA, Y ;
PITT, CG .
BIOMATERIALS, 1990, 11 (02) :108-112